Environmental samples

2

(d)

Sample collection devices can be a sponge hydrated with a neutralizing solution to

3

inactivate the effects of the sanitizers. 3M recommends the use of a biocide-free

4

cellulose sponge. Neutralizing solution can be Dey-Engley (D/E) Neutralizing Broth

5

or Letheen broth. It is recommended to sanitize the area after sampling.

6

7

WARNING: Should you select to use Neutralizing Buffer (NB) that contains aryl

8

sulfonate complex as the hydrating solution for the sponge, it is required to perform a

9

1:2 dilution (1-part sample into 1-part sterile enrichment broth) of the enriched

10

environmental sample before testing in order to reduce the risks associated with a

11

false-negative result leading to the release of contaminated product. Another option is

12

to transfer 10 µL of the NB enrichment in the LS tubes.

13

14

(e)

The recommended size of the sampling area to verify the presence or absence of the

15

pathogen on the surface is at least 100 cm

2

(10 cm x 10 cm or 4”x4”). When sampling

16

with a sponge, cover the entire area going in two directions (left to right then up and

17

down) or collect environmental samples following your current sampling protocol or

18

according to the FDA BAM, USDA FSIS MLG or ISO 18593 [11] guidelines.

19

1.

Allow the Demi-Fraser Broth enrichment medium (includes ferric ammonium

20

citrate) to equilibrate to ambient laboratory temperature (20-25

o

C).

21

2.

Aseptically combine the enrichment medium and sample according to Table

2A

.

22

3.

Homogenize thoroughly by

blendingvortexing or ,

stomaching

, or hand mixing

23

for 2 ±0.2 minutes. Incubate at 37 ±1°C for 24-30 hours.

24

25

E.

P

REPARATION OF THE

3M™M

OLECULAR

D

ETECTION

S

PEED

L

OADER

T

RAY

26

(a)

Wet a cloth or paper towel with a 1-5% (v:v in water) household bleach solution and

27

wipe the 3M™ Molecular Detection Speed Loader Tray.

28

(b)

Rinse the 3M Molecular Detection Speed Loader Tray with water.

29

(c)

Use a disposable towel to wipe the 3M Molecular Detection Speed Loader Tray dry.

30

(d)

Ensure the 3M Molecular Detection Speed Loader Tray is dry before use

31

32

Specific Instructions for Validated Methods

33

AOAC® Performance Tested Method

SM

#111501

34

35

36

37

Formatted:

Line spacing: single