Step 1:

Determine the uncertainty (

σ

) for the dose measurement,

yielding the

probability distribution for the true dose

.

Dm + C /2

Dm - C /2

Action limits - Proposed philosophy

Dose (Gy)

C

α

Dose measurement

= Dm

α

α

Error probability = (1-CL)/2

(i.e. 2.5% when CL=95%)

Step 2:

Set a confidence level CL for the true dose, e.g.

CL =95%, and determine the corresponding dose interval C

α

.

From Jörgen Olofsson (Dose Modeling and dose verification ESTRO COURSE)