Table of Contents Table of Contents
Previous Page  298 / 453 Next Page
Information
Show Menu
Previous Page 298 / 453 Next Page
Page Background

QIAGEN

mericon

STEC Workflow Collaborative Study Protocol

May 2016

DRAFT

8

C.

Apparatus

(

a

)

Incubator

.-Capable of maintaining 42 ± 1°C.

(

b

)

Top-loading balance.-

Capable of weighing 1-2000 g.

(

c

)

Laboratory paddle blender.-

Seward 3500 or equivalent.

(

d

)

Vortex mixer.-

Multiple laboratory suppliers (MLS).

(

e

)

Microcentrifuge.-

For 1.5-2.0 mL microcentrifuge tubes.

(

f

)

Pipettors.-

Adjustable, capable of delivering 10-100 μL and 100-1000 μL.

(

g

)

Dry bath incubator.-

Capable of maintaining 100 ± 1°C.

(

h

)

Calibrated thermometer.-

Capable of measuring 100 ± 1°C.

(

i

)

Thermocycler

.-RotorGene Q®, QIAGEN #9001640, with notebook computer containing Rotor-

Gene Q® software version 2.3.1.49 or later.

(

j

)

QIAsymphony® SP.-

QIAGEN #9001297, for automated DNA extraction.

(

k

)

QIAsymphony® AS.-

QIAGEN #9001301, for automated PCR assay setup.

(

l

)

Rotor.-

QIAGEN #9018903, 72 wells, for holding 0.1 mL strip tubes and caps.

(

m

)

Locking ring for 72 well rotor

.-QIAGEN #9018904, for locking strip tubes and caps in 72-well

rotor.

(

n

)

Rotor-Disc 72 Rotor.-

QIAGEN #9018899.

(

o

)

Locking ring for Rotor-Disc 72.-

QIAGEN #9018900.

(

p

)

Reagent cartridge holder (2).-

QIAGEN #997008.

(

q

)

Sample carrier insert, 2 mL v2

.-QIAGEN #9242083, secondary tube adapter for 2 mL screw-cap

tubes.

(

r

)

Cooling adapter.-

QIAsymphony EMT v2, QIAGEN #920730.

(

s

)

Sample preparation cartridges

.-8-well, QIAGEN #997002.

(

t

)

8-Rod covers

.-QIAGEN #997004.

(

u

)

Elution microtubes.-

CL with cap strips, QIAGEN #19588.

(

v

)

Reagent holder cooling adapter.-

QIAGEN #9018090.

(

w

)

Adapter.-

2 x Rotor-Disc, for use with QIAsymphony, QIAGEN #9242204.

(

x

)

Rotor-Disc 72 loading block.-

QIAGEN #9018910.Store refrigerated to ensure that PCR setup is

performed under stable thermal conditions.

(

y

)

Rotor-Disc 72

QIAGEN #981301 (24) or # 981303 (240)

(

z

)

Rotor-Disc heat sealing film

.-QIAGEN #981601 (60) or #981604 (600).

(

aa

)

Rotor-Disc heat sealer

.-QIAGEN #9018898 (110 V) or #9019725 (230 V).

(

bb

)

Filter tips.-

1500 μL, QIAGEN #997024.

(

cc

)

Filter tips.-

200 μL, QIAGEN #990332.

(

dd

)

Filter tips.-

50 μL, QIAGEN #997120.

D.

Additional Supplies, Reagents and Media

(

a

)

Filter laboratory blender bags.-

MLS.

(

b

)

Serological pipettes –

Aerosol resistant

(

c

)

Micropipette tips,

aerosol resistant.-MLS.

OMAMAN-36 B : Collaborative Study Protocol

For ERP Use Only

January 2017

AOAC Research Institute