35)After the run is finished, select “Remove” in the assaysetup “Overview” screen.
Open the “Assays” drawer and unload the Rotor-Disc 72.
36)Place the Rotor-Disc sealing film over the Rotor-Disc72 and transfer to the
Rotor-Disc heat sealer. Sealing is complete in about 2 minutes. Remove the
excess film from the Rotor-Disc 72 and transfer to the Rotor-Disc 72
Rotor.Inthe
“Sample Rack Layout” screen of the assay setup user interface, the elution rack
in slot 2 is pictured.
Analysis
(a) Prepare Equipment
1) Turn on the Rotor-Gene Q instrument and attached computer. Load the
Rotor-Gene Q Series Software and templates for the
mericon
E. coli O157
Screen Plus and
mericon
E.coli STEC O-type assays. Start an empty run to
warm up the Rotor-Gene Q.
i. Run a Disc
ii. Load Samples
2) Manual mericonDNA Extraction Method
i. Load the QIAGEN strip tubes into the 72-Well Rotor strip tube ring.
Snap the locking ring in place over the strip tubes, and place into
the Rotor-Gene Q. When running strip tubes, balance the ring
using the empty strip tubes.
3) Automated QIAsymphonyDNA Extraction Method
i. Load the QIAGEN 72 Rotor-Disk into the 72-Rotor Disk ring. Snap
the locking ring in place over the Rotor Disk, and place into the
Rotor-Gene Q.
4) Start Instrument Run
i. Software Version 2.3.1 (Build 49)
Follow the Rotor-Gene software instructions. Verify that the correct
setting is selected for each preparation method: the Rotor-Disc 72
rotor is selected for the QIAsymphony (automated preparation
method), and the 72 count strip tubes for the manual preparation
method. Ensure that the locking ring is in place and select “Next”.
Verify that the reaction volume is 20 µL and select “Next”. Select
“Next” after verifying the temperature profile. Select “Start Run” to
start the Rotor-Gene Q instrument.