3) Prepare the reconstituted assay by adding 1040 µL(96 sample kit) or 130 µl (24
sample kit) of the Multiplex PCR master mix to a vial of
mericon
Assay and mix
by pipetting up and down 2-3 times. Prepare a positive control by re-suspending
the lyophilized Positive Control DNA in 200µL of QuantiTectNucleic Acid Dilution
Buffer and mix. Prepare the appropriate number of strip tubes by transferring
10µL of the reconstituted assay to the reaction tubes. Pipette 10µL of the
extracted sample DNA into each sample strip tube and mix by pipetting up and
down 2-3 times. Prepare the positive control tube by pipetting 10µL of the
Positive Control DNA into a strip tube containing 10 µL of reconstituted assay.
Prepare the negative control by pipetting 10 µL of RNase-free water into a strip
tube containing 10 µL of thereconstituted assay. Seal the strip tubes with the
strip tube caps. Load the samples into the Rotor-Gene-Q and initiate the analysis
immediately.
(b) Automated QIAsymphonyDNA Extraction Method
1) Ensure the “Waste” drawer is prepared properly, and perform an inventory scan
of the “Waste” drawer, including the tip chute and liquid waste. Replace the tip
disposal bag, if necessary.
2) Load an elution microtube rack (EMTR) into the “Eluate” drawer and perform an
inventory scan of the “Eluate” drawer.
3) Load the
QIAsymphonymericon
bacteria kit and consumables into the “Reagents
and Consumables” drawer.Select the “R+C” button on the touchscreen to open
the screen that shows the consumables status (“Consumables/8-
RodCovers/Tubes/ Filter-Tips/Reagent Cartridges”).Select the “Scan Bottle”
button to scan the bar code of the TopElute bottle with the handheld bar code
scanner. Select the “OK” button.
4) Perform an inventory scan of the “Reagents andConsumables” drawer.
5) Transfer 500 µL of the enriched sample into a 2 mL screw cap tube.Place the
samples into the appropriate tube carrier and load them into the “Sample”
drawer.
6) Using the touchscreen, enter the required information for each batch of samples
to be processed.
7) Choose elution volumes according to Table 1.
Table 1: Matrix Manual Dilution and Automated Elution Volumes
Matrix
Manual
mericon
Method
DNA Dilution
Automated
QIAsymphony
Method
DNA Elution Volume
Fresh Spinach
1:10
400 µL
OMAMAN-36 E/ AOAC PTM Report
ERP Use Only
January 2017