Dual Lab Validation of a Method for the Determination of Fructans in Infant 

Formula & Adult Nutritionals /        

25 Jul 2016

16 

3 R E S U L T S 

3.1.

System Calibration 

A quadratic model was used to fit the data for calibration purposes. The relative residuals were 

calculated and plotted against the analyte concentration.  For glucose the quadratic line fits the data 

well (Figure 1), and the predicted concentration is never more than 2.2% away from the actual 

concentration (Figure 2). 

Figure 1: Glucose Calibration Curve

Figure 2: Plot of Relative Residuals (%) Against Glucose Concentration

For fructose, the quadratic line fits the data well (Figure 3), the predicted concentration is always 

within 2% of the actual concentration except at the lowest concentration (20 µg/mL) where there 

can be up to a 6 % difference (Figure 4). 

-6.00

-4.00

-2.00

0.00

2.00

4.00

6.00

0

50

100

150

200

250

300

350

Differnce [%]

Glucose conc. [µg/mL]

GLUCOSE: Plot of the differences vs concentrations

VALIDATION REPORT