Table 2. Residuals and error of back-calculated concentrations for all-
trans
-lutein standard
curves.
Experiment 1
Experiment 2
Experiment 3
Conc
(μg/100 mL)
Residual
Error
Residual
Error
Residual
Error
235.9
0.000368916 0.01% -0.000497233 -0.02% -6.16301E-05 0.00%
125.8
-0.003210486 0.24%
-0.00122698 0.09%
0.00221285 -0.16%
78.6
0.00379152 -0.44% 0.
003308385 -0.38% -0
.004609309 0.54%
31.5
0.000817163 -0.24% 0.000644834 -0.18% 0.003230669 -0.93%
3.93
-0.
001767113 3.78
% -0.002229005 4.65%
-0.00077258 1.65%
Figure 2. Linearity plot for lutein.
Table 3. Residuals and error of back-calculated concentrations for all-
trans
-β-carotene
standard curves.
Experiment 1
Experiment 2
Experiment 3
Conc
(μg/100 mL)
Residual
Error
Residual
Error
Residual
Error
227.4
0.00125077 0.05%
0.002290766 0.09% -0
.001591824 -0.06
%
121.3
-0.002520458 -0.18% -0.003680882 -0.27%
0.00318571 0.23%
75.8
-0.000366823 -0.04%
-0
.00193471 -0.23% -0
.000338007 -0.04%
30.3
0
.001615483 0.47
%
0.002244522 0.65%
0.000226108 0.06%
3.79
2.10278E-05 0.04%
0
.001080305 2.33%
-0
.001481986 -2.97%
VALIDATION REPORT
FOR ERP USE ONLY
DO NOT DISTRIBUTE