Table of Contents Table of Contents
Previous Page  88 / 220 Next Page
Information
Show Menu
Previous Page 88 / 220 Next Page
Page Background

Figure 10. SPIFAN Toddler formula.

Figure 11. SPIFAN High protein adult nutritional RTF.

Precision

All samples were prepared in triplicate and the SPIFAN samples with substantial carotenoid

concentrations were prepared a at least 12 times (3 replicates x 4 days). Intermediate

precision (RSD

iR

) was calculated as the relative standard deviation of all replicates.

Repeatability (RSD

r

) for triplicate samples was calculated as the relative standard deviation,

and for larger samples sizes it was estimated by performing a one-way ANOVA on the

intermediate precision data and calculating the standard deviation from the within-day

variance. Although the SMPR does not specify criteria for intermediate precision, the

requirements for repeatability and reproducibility are outlined in Table 6. For all

measurements above 1 μg/100 g, the method meets the precision requirements of the SMPR.

Table 6. Precision requirements from SMPR 2014.014.

Sample conc (μg/100 g)

Repeatability

Reproducibility

1-100

8%

15%

>100-1300

5%

10%

VALIDATION REPORT

FOR ERP USE ONLY

DO NOT DISTRIBUTE