Figure 10. SPIFAN Toddler formula.
Figure 11. SPIFAN High protein adult nutritional RTF.
Precision
All samples were prepared in triplicate and the SPIFAN samples with substantial carotenoid
concentrations were prepared a at least 12 times (3 replicates x 4 days). Intermediate
precision (RSD
iR
) was calculated as the relative standard deviation of all replicates.
Repeatability (RSD
r
) for triplicate samples was calculated as the relative standard deviation,
and for larger samples sizes it was estimated by performing a one-way ANOVA on the
intermediate precision data and calculating the standard deviation from the within-day
variance. Although the SMPR does not specify criteria for intermediate precision, the
requirements for repeatability and reproducibility are outlined in Table 6. For all
measurements above 1 μg/100 g, the method meets the precision requirements of the SMPR.
Table 6. Precision requirements from SMPR 2014.014.
Sample conc (μg/100 g)
Repeatability
Reproducibility
1-100
8%
15%
>100-1300
5%
10%