Linearity

Linearity of the relative responses of analyte concentrations was measured using a 5-point

standard curve on three different days. Correlation coefficients, visual inspection, residuals,

and relative errors of back-calculated concentrations were used for evaluation. Linearity of

the internal standard was also demonstrated.

The ranges for lutein and β-carotene (4-240 μg/100 mL) correlate to approximately 0.8-45

μg/100 g as fed for samples prepared for the lowest sample concentrations. With dilutions

specified in the method, the range can be extended to approximately 2250 μg/100 as fed. This

extends beyond the range of 1-1300 μg/100 g specified in the SMPR.

Table 1. Residuals for the internal standard

Apocarotenal

Conc

(μg/100 mL)

Residual

203.5

4.119148067

101.8

-7.399430255

40.7

-3.177165248

20.4

1.435593088

4.07

2.222225757

2.04

2.79962859

Figure 1. Linearity plot for apocarotenal.