

Pacquette & Thompson:
J
ournal of
AOAC I
nternational
V
ol.
98, N
o.
6, 2015
1703
were equipped with modern collision/reaction cells that are
thought to be necessary to avoid the Ar/C spectral interferences
on the major Cr and Se isotopes.
Before actual MLT study samples were analyzed, each
participating laboratory was asked to set up the method and
evaluate the linearity and the method LOQ with their given
instrument model. This exercise is identical to what is done to
transfer a mineral method to another site in the authors’ internal
laboratory network, as it quickly identifies problems in procuring
or preparing suitable standards and standardblanks, or inotherwise
setting up the instrument parameters. To check the linearity,
standards were analyzed and the calibration curve prepared on
each of 3 separate days. On each day, working standards at the
lowest concentration level (WS1) and at ½WS1 were analyzed as
samples, and then their calculated concentrations were compared
to their nominal concentrations. The mean recovery of each
standard versus its nominal concentration (i.e., the calibration
residual) had to be within 5%. All laboratories passed this test
except Laboratories 9 and 11, both of which failed only at the
lowest standard level for Se (Table 1). For these laboratories,
the practical LOQ (PLOQ) for Se was therefore equal to WS1,
whereas the other laboratories could analyze as low as ½ WS1
in concentration.
The second setup test was to analyze the sample blank on
5 separate days (done at same time as the linearity study, plus
two more days), calculating the SD of these results, multiplying
that by 10, and then adding that result to the blank mean. This
LOQ was multiplied by the method’s dilution factor of 50 to
arrive at the approximate LOQ in terms of sample weight. The
SMPRs state an LOQ of 20 ng/g Cr and Mo and 10 ng/g Se on
a ready-to-feed (RTF) basis. Table 1 shows the prework results
from the participating laboratories (Laboratories 6 and 7 dropped
out about this time). Note that it is desirable to have low,
consistent blanks for good sensitivity, as well as the linearity, to
avoid excessive calibration bias. These trials immediately pointed
to Laboratories 1 and 10 as having potential problems; they were
allowed to proceed with the MLT, but indeed Laboratory 1’s
data were eventually rejected in total. The prework results for
Laboratory 10 may not have been so ominous because it did not
submit all the data, and the PLOQs could not be calculated.
The final prework for the participating laboratories was to
analyze the NIST Standard Reference Material (SRM) 1849a
sample. All laboratories passed this test by producing Cr, Mo, and
Se results within 5% of the certified means (data not shown, but
similar to data collected during the MLT, which is tabulated later).
The fact that Laboratories 1 and 10 produced good results on the
SRM might be attributed to the relatively high concentration of
these elements in the SRM. It should be noted that six of the nine
laboratories determined Na, K, P, Mg, Ca, Fe, Cu, Zn, and Mn
concurrently with the Cr, Mo, and Se with good precision and
accuracy. This work was done under the direction of the Study
Director. Results on SPIFAN samples are published in this
issue of
J. AOAC Int.
[AOAC First Action Method
2015.06
by
Thompson, J.J., Pacquette, L., & Brunelle, S.L. (2015)
J. AOAC
Int.
98
, 1711–1720]. Method
2011.19
appears to be viable as a
12-element method, not just as a method for ultratrace elements.
Each participating laboratory received blind duplicates of
seven of the SPIFAN matrixes (this study used the original
SPIFAN set) for a total of 14 samples to test. NIST SRM 1849a
was not included as a blind sample, but rather the participants
were instructed to analyze it concurrently with the other samples
as if it were a control sample. The seven matrixes tested were an
infant formula partially hydrolyzed milk-based powder, an adult
nutritional low-fat powder, an adult nutritional milk protein-
based powder, a child formula powder, an infant elemental
powder, an adult high protein nutritional RTF liquid, and an adult
high-fat nutritional RTF liquid. Only two infant formula types
were chosen (there were four more in the SPIFAN set) because
they were known to be unfortified in Cr and Mo and would not
yield useful information.
Participants were asked to reconstitute all powders prior to
analysis with the exception of SRM 1849a, which was unblinded
but rather easily identified by its sachet anyway. Participants used
a direct weight of 0.2 g SRM powder, which has proven to be
homogeneous for minerals at this weight through extensive use in
the authors’ laboratories. All other powders were reconstituted by
either dissolving 20 g powder in enough laboratory water to make
200 g solution, i.e., a 10% (w/w) reconstitution, or by following
the official method with the SPIFAN-recommended 25 g sample
+ 200 g water (11.1%, w/w). Some laboratories asked to work
with the 10% reconstitution rates, as this is certainly an easier
Table 1. Set-up tests for participating laboratories
a
Lab
LOQ Cr (20 ng/g
required), ng/g
PLOQ Cr, µg/L
LOQ Mo (20 ng/g
required), ng/g
PLOQ Mo, µg/L
LOQ Se (10 ng/g
required), ng/g
PLOQ Se, µg/L
1
45
0.4
30
0.4
46
0.2
2
7
0.4
5
0.4
4
0.2
3
9
0.4
9
0.4
3
0.2
4
12
0.4
2
0.4
1
0.2
5
9
0.4
1
0.4
6
0.2
6
7
8
4
0.4
1
0.4
1
0.2
9
16
0.4
13
0.4
18
0.4
10
44
?
14
?
66
?
11
8
0.4
4
0.4
13
0.4
a
Laboratories 6 and 7 dropped out at this time and Laboratory 10’s data were incomplete. A PLOQ of 0.4 µg/L Cr/Mo and 0.2 µg/L Se along with an
LOQ below 20 ng/g (10 ng/g for Se), was desired.
See
text for details.
158