qualitative

cis/trans

C18:1 FAME isomers standard mixture

solution,

D

(

k

). The resolution is sufficient when resolution (R)

criteria is equivalent or higher than 1.00 (Figure

2012.13C

).

The following two examples report applicable conditions for

a correct separation of

cis

and

trans

with different GC injectors.

(b) 

Example 1.

—Split injection mode.

(

1

) 

Column.

—100 m length × 0.25 mm id, 0.2 μm film

thickness, fused silica capillary column.

(

2

) 

Stationary

phase

.—Cyanopropyl-polysiloxane

or

equivalent polarity.

(

3

) 

Carrier gas type.

—Helium.

(

4

) 

Column head carrier gas pressure.

—225 KPa

(175–225 KPa).

(

5

) 

Split flow

.—25.5 mL/min.

(

6

) 

Split ratio.

—10:1.

(

7

) 

Injector temperature.

—250°C.

(

8

) 

Detector temperature.

—275°C.

(

9

) 

Oven temperature program.

—Initial temperature

of 60°C, maintained for 5 min, raised at a rate of 15°C/min up

to 165°C, maintained at this temperature for 1 min, and then

raised at a rate of 2°C/min up to 225°C for 20 min.

(

10

) 

Amount of sample injected.

—1.0 μL.

An example of the GC profile obtained with these conditions

is reported in Figure

2012.13A

.

(c) 

Example 2.—

On-column injection mode.

(

1

) 

Column

.—100 m length × 0.25 mm id, 0.2 μm film

thickness, fused silica capillary column.

(

2

) 

Stationary

phase

.—Cyanopropyl-polysiloxane

or

equivalent polarity.

(

3

) 

Carrier gas type.

—Hydrogen.

(

4

) 

Column head carrier gas pressure.

—210 KPa

(175–225 KPa).

(

5

) 

Injector temperature.

—Cold.

(

6

) 

Detector temperature.

—275°C.

(

7

) 

Oven temperature program

.—Initial temperature

of 60°C, maintained for 5 min, raised at a rate of 15°C/min up

to 165°C, maintained at this temperature for 1 min, and then

raised at a rate of 2°C/min up to 225°C for 17 min.

(

8

) 

Amount of sample injected.

—1.0 μL.

An example of the GC profile obtained with these conditions

is reported in Figure

2012.13B

.

(d) 

Resolution between C18:1

trans

and

cis

.

—Inject into the

gas chromatograph 1.0 μL calibrating solution,

D

(

k

). Determine

peak width at half height and distance between the left of the

Figure 2012.13C. Example of a GC chromatogram (i.e., with insufficient and sufficient resolution for C18:1

trans

).