FOR EXPERT REVIEW PANEL USE ONLY

NOT FOR DISTRIBUTION

2

Method is straightforward, without complex sample preparation

Widely available equipment is used

Considerable reduction in analysis time

Cons/Weaknesses

Previous remarks were taken up satisfactorily

Supporting Data

General Comment:

Wide range of matrices tested for 3 types of fructans, including SPIFAN samples

Multiple labs, analysts, and equipment used

Performance Characteristics:

Analytical Range: 0.03 – 5 % as-fed, 45-100% ingredient commodity

LOQ: 0.03 % (empirically determined using low spike level in sugar solution)

Accuracy/Recovery: average recoveries within 90 – 110% range

Precision (RSD

r

): well within acceptability limits

Reproducibility (RSD

R

): well within acceptability limits

System suitability: ok

Recommendation:

Clear method with extensive supporting validation data. We recommend this method for First Action

approval.

FOS-04 Review Forms