Table of Contents Table of Contents
Previous Page  130 / 159 Next Page
Information
Show Menu
Previous Page 130 / 159 Next Page
Page Background

FOR EXPERT REVIEW PANEL USE ONLY

NOT FOR DISTRIBUTION

3

A quadratic curve has been used for calibration. Only 4 points are employed for making the quadratic

curve (may be the limit for a quadratic) – plot of relative error indicates in most cases the error is less

than 5% (one exception at lowest level).

Performance Characteristics:

Analytical Range: tested

0.05 – 3.0 (SMPR goes up to 5.0)

LOQ: assessed to be 0.03 g/100g in water solution containing sucrose. Might also be interesting to

assess in blank matrix.

Accuracy/Recovery:

92.9 – 108%

(meets SMPR 90 – 110).

Precision (RSD

r

):

1.09 – 3.67

(meets SMPR <= 6)

(RSD iR): 2.46 – 6.79 (no specification in SMPR)

Reproducibility (RSD

R

): n/a

System suitability:

-

Blank samples have been analysed and found to be blank.

-

Otherwise no mention of the use of check standards or blank check samples

Recommendation:

- Write method in appropriate format

- Better describe the actual spiking levels in spike/recovery experiments

- Good candidate for 1

st

action

FOS-04 Review Forms