FOR EXPERT REVIEW PANEL USE ONLY

NOT FOR DISTRIBUTION

2

Data for alpha-carotene and Lycopene were not collected

Method Clarity:

Method clearly written and should be easy to follow.

Method Safety Concerns:

No safety concerns are noted by the authors

Pros/Strengths:

Well-established methodology

Clearly written and easy to follow

Good chromatography for listed carotenoids

Repeatability and recovery data for lutein and β-carotene meet SMPR

Acceptable linearity

Cons/Weaknesses

Not all carotenoids specified by SMPR are determined by this method. The authors are provided the

following explanation for exclusion of a-carotene and lycopene: “While alpha‐carotene is separated

with this method, it was not included in the determination because only one SPIFAN matrix contained

the analyte. Lycopene is also separated, but is excluded due to poor standard injection precision and

poor standard stability.

Even though validation data are not collected for α-carotene and lycopene both compounds were shown to

be separated by the present method from all lutein and β-carotene isomers

Supporting Data

General Comment:

-

Method Optimization:

Performance Characteristics: