Analytical Range:

0.8-2250 ug/100 g – meets SMPR

LOQ:

For both lutein and β-carotene 0.27 ug/100 g (meets SMPR)

Accuracy/Recovery:

Accuracy by spike recovery at 2 levels of all-trans lutein and all-trans b-carotene were completed

for

of SRM1849a as well as 2 other matrices – meet SMPR for recoveries and repeatability (no

reference or certified carotenoid values available from NIST)

Precision (RSD

r

):

All SPIFAN matrices were analyzed and repeatability and intermediate reproducibility meet SMPR.

Only few matrices had repeatability higher than specified by SMPR abut these matrices had the analytes

present at concentrations below the analytical range specified by SMPR

Reproducibility (RSD

R

):

no data available

System suitability:

System suitability parameters are clearly defined. The method specifies how to

prepare system suitability mixtures of isomers and what resolution between them is expected

Recommendation:

I recommend this method as AOAC First Action method for

determination of lutein and β-carotene only in

infant formula and adult nutritionals providing the scope of the method is clearly defined.