2. Is there information

demonstrating that the

method meets the SMPR

Method Performance

Requirements using the

Reference Materials stated in

the SMPR? If not, then

specify what is missing and

how this impacts

demonstration of

performance of the method.

3. Is there information

demonstrating that the

method performs within the

SMPR Method Performance

REquirements table

specifications for all analytes

in the SMPR applicability

statement? If not, please

specify what is missing and

whether or not the method's

applicability should be

modified.

1.

Based on the supporting

information, were there any

additional steps in the evaluation of

the method that

indicated the need

for any additional precautionary

statements in the method?

2. Is there information

demonstrating that the

method system suitability

tests and controls as

specified in the SMPR worked

appropriately and as

expected? If no, please

specify.

3. Based on the supporting

information, is the method

written clearly and concisely?

If no, please specify the

needed revisions.

4. Based on the supporting

information, what are the

pros/strengths of the

method?

YES

YES

YES, Internal standards should be added prior to the extraction to understand the

actual recoveries of analytes in sample matrices.

YES

YES however SRM results must be reported.

1.Most confirmative method as it is MRM method with eight transitions per allergen.

2. Used SRMs and labeled standards

3. Linearity studies conducted as per SMPR.

IV. General Submission Package: