Zero Margins in daily online ART/IGRT?

Intrafraction motion: prostate

A. N. T. J. Kotte et al. 2007: Intrafraction motion of the prostate during external-beam

radiation therapy: Analysis of 427 patients with implanted fiducial markers. IJROBP 69(2)

Motion of implanted fiducial markers as seen with PIs:

11426 fractions of 427 pts.

~5-7min after start of

treatment

ESTRO IMRT course 2016

Adaptive Radiotherapy

London April 7th, 2016

-34-

intrafraction motion range >2mm in 66% of all fractions, >3mm in 28%