Table of Contents Table of Contents
Previous Page  423 / 1082 Next Page
Information
Show Menu
Previous Page 423 / 1082 Next Page
Page Background

S408

ESTRO 36 2017

_______________________________________________________________________________________________

Conclusion

All together these results strongly support the fact that an

accurate dosimetry needs to be performed before an

experiment but also to cautiously follow all the defined

parameters for one condition of irradiation to avoid errors

in the dose delivered on the sample and to be able to

properly compare and interpret experiments.

PO-0778 New Razor silicon diode for Cyber Knife small

beam relative dosimetry: a multi-site evaluation

S. Russo

1

, L. Masi

2

, P.R. Dicarolo

3

, R. Doro

2

, E. De

Martin

4

, M.L. Fumagalli

4

, A.S. Martinotti

5

, A. Bergantin

5

,

E. Rondi

6

, S. Vigorito

6

, P. Mancosu

7

1

Aziend USL Toscana Centro, Fisica Sanitaria, Florence,

Italy

2

IFCA, Radiotherapy, Firenze, Italy

3

Medical Physycs, Meyer Children’s University Hospital-,

Florence, Italy

4

Istituto Besta, Radiotherapy, Milano, Italy

5

C.D.I., Radiotherapy, Milano, Italy

6

I. E. O., Radiotherapy, Milano, Italy

7

Humanitas Research Hospital, Medical Physics Unit of

Radiation Oncology, Milan, Italy

Purpose or Objective

The aim of this work was to evaluate the suitability of a

new unshielded p-type silicon diode (Razor, IBA

Dosimetry, Germany) for relative small beams dosimetry

over different CyberKnife systems.

Material and Methods

Output Factors

(OFs) measurements with Razor detector

were performed by four Italian Radiotherapy Centers

equipped with CyberKnife units for field sizes ranging from

5 to 60 mm, defined by fixed circular collimators. Setup

conditions were 80 cm source to detector distance and 1.5

cm depth in water. Measurements were repeated by each

center with a PTW-60017 diode. Monte Carlo correction

factors reported in literature were applied to PTW-60017

measured data and corrected values were considered as a

reference.

Crossplane and inplane dose profiles ranging from 5-60 cm

fixed collimators were measured by Razor detector at a

depth of 10 cm in water and SSD 70 cm. The effective field

size (EFS), defined as EFS=, where A and B correspond to

the in- and cross-line FWHM, were calculated. Penumbra

20%- 80% was also evaluated.

This work has been conducted in the framework of the

Italian Association of Medical Physics (AIFM) SBRT working

group.

Results

Razor OFs measured for fixed collimators in the four

enrolled centers showed a variability (relative range)

decreasing from 1.2% to 0.4% for field sizes from 7.5 to 60

mm and equal to 2.2% for the smallest cone. The

variability obtained for OF measured by PTW-60017 was

analogous: lower than 1 % for field sizes from 7.5 to 60

mm and equal to 3.5% for the smallest diameter.

For field sizes down to 7.5 mm Razor measured OFs were

lower than PTW-60017 uncorrected measured values.

Relative differences between Razor OFs and Monte Carlo

corrected PTW-60017 data were below 1% for 60-10 mm

cone sizes and within 2 % for 7.5 mm field size over all

centers. For the smallest collimator differences ranging

from to 2.5% to 6% were observed among centers. Average

values and SD of OFs measured by Razor and PTW-60017

diode (MC corrected and not) are shown in figure.

Nominal field size NFS, effective field size EFS and

penumbra Razor measurements averaged over the four

CyberKnife centers are reported in table. Maximum

difference between NFS and EFS was about 6% for 5 mm

field size. Penumbra values were lower than 3 mm for field

sizes up to 15 mm.

Conclusion

Conclusions

: CyberKnife OFs measured by Razor showed a

high consistency among different centers and a

comparable variability to data obtained by PTW-60017

routine detector. Comparison between Razor OFs and

PTW-60017 measurements corrected by Monte Carlo

indicated that correction factors for Razor should be

smaller than for PTW-60017 down to 7.5 mm field size. EFS

and penumbra measured over the four centers showed a

good consistency confirming Razor as a good candidate for

small beam relative dosimetry.

PO-0779 New robotic phantom for evaluation of

imaging and radiotherapy of moving structures

H. Arenbeck

1

, L. Eichert

1

, G. Hürtgen

2

, K. Gester

2

, I.

Brück

2

, N. Escobar-Corral

2

, M. Fleckenstein

1

, A. Stahl

3

,

M.J. Eble

2

1

Boll Automation GmbH, Research and Development,

Kleinwallstadt, Germany

2

RWTH Aachen University Hospital, Radiooncology and

Radiotherapy, Aachen, Germany

3

RWTH Aachen University, III. Institute of Physics B,

Aachen, Germany

Purpose or Objective

Four dimensional radiotherapy processes that allow an

adaptation to intrafractional motion require increased

accuracy of dose application while displaying increased

technological and procedural complexity and thus

multiplied sources of error. Consequentially, novel 4D

phantoms are required that feature anthropomorphic

structure and motion. In this work, a prototype of such

phantom, which is fit for long term clinical service, is

presented.

Material and Methods

The modular phantom architecture allows different static

and moving human equivalent structures and dose

measurement devices to be placed into the irradiated

region. A new kind of parallel robot generates freely

programmable motion in all Cartesian directions. The

whole system is portable and features similar extension as

a human. Concept, kinematics, construction and software

of a previously presented evaluation model have been

fundamentally refined.