CHROMATOGRAPHY

Multimode ODS column:

Scherzo SS-C18 column (100 x 3 mm)*

Purified Porous Silica / Particle Size 3um / Pore Size 13nm / ODS

+ Anion Exchange + Cation Exchange Ligands

Separation achieved through a manipulation of eluent pH, organic

solvent composition, salt concentration, and temperature.

Enables the retention of ionic species with LC-MS compatible

mobile phases

Avoids the use of ion pairing agents in mobile phases

Proprietaryand confidential — do not distribute

© 2015 Abbott Laboratories

Provides simple and rugged chromatographic retention

*= Imtakt USA, Portland, OR

March17,2015

MFA‐AOAC, 2015 

11

SAMPLE PREPARATION

For powders:

Weigh 10.0

0.5 g of powder into a tared vessel and record mass.

Add water to a total weight of 70

5 g (powder + water) and record

mass.

Stir until homogenous.

For ready-to-feed liquids and reconstituted powders:

Weigh 5.00

0.20 g liquid into a tared 15 mL centrifuge tube and

record mass.

Add 40

L of internal standard working solution and vortex.

Add 50 μL of glacial acetic acid and 5.0 mL methanol, vortex.

Transfer aliquot to microfuge tube and centrifuge at 13,000 rpm for

Proprietaryand confidential — do not distribute

© 2015 Abbott Laboratories

10 min.

Filter an aliquot of the supernatant into an autosampler vial using a

0.2 um PTFE syringe filter.

March17,2015

MFA‐AOAC, 2015 

12