Background Image
Table of Contents Table of Contents
Previous Page  62 / 116 Next Page
Information
Show Menu
Previous Page 62 / 116 Next Page
Page Background

ANTH-002

System Suitability

Not stated.

Safety Review

Is the method safe to use as written?

Yes, if the user is a trained chemist, and has working

knowledge of handling chemicals and instrumentation.

Are there any safety concerns with

any component, instrument, reagent

or instruction of the method? If yes,

please clearly state the concerns.

There is no safety section in the method. The method

doesn't have any statement/warning about the safe

handling of chemicals used in the method.

Does the method include all

necessary safety precautions and/or

warning statements? Yes or No. If

No, please suggest potential

precautions or warnings that can be

added to the method for users or

recommend safer option(s) if

warranted.

No

Recommendations

Final Recommendation:

This method doesn't satisfy the performance

requirements stated in the SMPR; it's a quantitative

method for the determination of total anthocyanins and

not an authentication method. Not recommended for

further consideration.