ANTH-002

analysis or different processing methods used to

calculate both standards and samples together at

different wavelengths to ensure consistency.

Clarity required around standard concentration

calculation. The standard listed in method is salt form.

No discussion of handling/drying of standard material

covered. Results are then Cyanidin-3-chloride, or

converted to Cyanidin-3-glucoside with correction factor

of 1.38 as listed in 8.6.4. This factor is incorrect, it should

be 1.3926, molecular mass data from Chem

Spider:(Cyanidin-3-glucoside 449.384 Da; Cyanidin

chloride 322.697 Da)

No QC, acceptance criteria or performance criteria in the

method. Either a NIST or other validated control needs

to be included.

No expiration dates on solutions, standards or samples,

or storage conditions of such. The filter type used to

filter samples is not stated.

There is no discussion of light sensitivity of the analytes

and working under yellow lights or using actinic

glassware.

There is no information in regard to the ruggedness of

the method when dealing with dietary supplement

matrices.

Comparing "fingerprints" of analytes may be difficult

between different HPLC systems/conditions as well as

growing conditions/harvest times. This could be

elucidated with example chromatograms.

Supporting Data

General Comments

No example chromatograms are included, so cannot

comment on resolution of peaks of interest or the

sensitivity.

Method Optimization

Sonication of samples and/or other extraction solvents

might be explored for increased extraction efficiency of

analytes.

Performance Characteristics

Analytical Range:

18 to 180 mg/L standard curve range.