where R = results expressed in mg/100g, C = concentration of the analyte in the injected solution in

ng/mL, V = volume of the initial extract in mL, S = sample weight in g, and D = dilution factor, the inverse

of any dilution made. All results were calculated on a ready to feed or reconstituted basis of 25 g

diluted to 225 g with water, except for SRM1849a that was calculated back to powder basis.

References

J. AOAC Int

. (future issue)

AOAC SMPR 2012.010

J. AOAC Int.

96

, 488(2013)

DOI: 10.5740/jaoac.int.SMPR2012.010

AOAC SMPR 2012.013

J.AOAC Int.

96

, 492(2013)

DOI: 10.5740/jaoac.int.SMPR2012.013

Posted: October 1, 2015

Table 2015.10A. MS/MS Settings

Ionization

Positive Ion Electrospray (ESI+)

Mode

MRM

IonSpray Voltage

1000 V

TurboIon Spray Temp

550°C

Declustering Potential

120 V

Dwell Time

100 msec

Entrance Potential

10 V

Collision Cell Exit Potential 25 V

Collision Gas

Nitrogen, 5 psig

Curtain Gas

Nitrogen, 20 psig

Nebulizing Gas (Gas 1)

Nitrogen, 60 psig

Auxiliary Gas (Gas 2)

Nitrogen, 60 psig

Needle Position

Y = 5mm, X = 5 mm