where R = results expressed in mg/100g, C = concentration of the analyte in the injected solution in
ng/mL, V = volume of the initial extract in mL, S = sample weight in g, and D = dilution factor, the inverse
of any dilution made. All results were calculated on a ready to feed or reconstituted basis of 25 g
diluted to 225 g with water, except for SRM1849a that was calculated back to powder basis.
References
J. AOAC Int
. (future issue)
AOAC SMPR 2012.010
J. AOAC Int.
96
, 488(2013)
DOI: 10.5740/jaoac.int.SMPR2012.010
AOAC SMPR 2012.013
J.AOAC Int.
96
, 492(2013)
DOI: 10.5740/jaoac.int.SMPR2012.013
Posted: October 1, 2015
Table 2015.10A. MS/MS Settings
Ionization
Positive Ion Electrospray (ESI+)
Mode
MRM
IonSpray Voltage
1000 V
TurboIon Spray Temp
550°C
Declustering Potential
120 V
Dwell Time
100 msec
Entrance Potential
10 V
Collision Cell Exit Potential 25 V
Collision Gas
Nitrogen, 5 psig
Curtain Gas
Nitrogen, 20 psig
Nebulizing Gas (Gas 1)
Nitrogen, 60 psig
Auxiliary Gas (Gas 2)
Nitrogen, 60 psig
Needle Position
Y = 5mm, X = 5 mm
Candidates for 2016 Method of the Year
121