: Nonspecific binding can occur with these analytes when using glassware, so plasticware should be

used at all times for standard/sample preparation. All laboratory plasticware should be single-use

whenever possible. Positive displacement pipets are also mandatory for pipeting to avoid

contamination and for accuracy with organic solvents.

B.

Chemicals and Reagents

(a)

Water

. – Optima MS grade (Thermo Fisher Scientific, Waltham, MA) or equivalent.

(b)

Acetonitrile

. – Optima MS grade (Thermo Fisher Scientific) or equivalent.

(c)

Ammonium formate

. – Optima MS grade (Thermo Fisher Scientific) or equivalent.

(d)

Formic acid

. – Optima MS grade (Thermo Fisher Scientific) or equivalent.

(e)

Nitric acid

. – 70% w/w, ACS grade (Avantor, Center Valley, PA) or equivalent.

(f)

Isopropanol

. – Optima MS grade (Thermo Fisher Scientific) or equivalent.

(g)

Desiccant

. – (VWR) or equivalent.

(h)

Reference Standard

. – L-Carnitine (USP, Rockville, MD) or equivalent.

(i)

Reference Standard

. – Choline Bitartrate (TCI, Tokyo, Japan) or equivalent.

(j)

Reference Internal Standard

. – L-Carnitine-d3 HCl (CDN Isotopes, Pointe Claire, Québec, Canada) or

equivalent.

(k)

Reference Internal Standard

. – Choline-1,1,2,2-d4 Chloride (CDN Isotopes) or equivalent.

Note

: All use of water in this method must be high purity MS grade water. It is recommended that all

preparation steps with nitric acid be performed within a fume hood, and the necessary personal

protective equipment used when handling.

C.

Mobile Phase Preparation

Mobile phase A (5mMammonium formate in 50:50 (v/v) water:acetonitrile with 0.2% formic acid) was

prepared by weighing 0.63 g of ammonium formate into a 1 L graduated cylinder. Water was added

along with a stir bar and mixed to dissolve before diluting to volume with water. The solution was

transferred to a 2 L mobile phase container along with 1 L of acetonitrile, 4mL of formic acid, stir bar,

and then thoroughly mixed. Mobile phase B (30mMammonium formate in 50:50 (v/v)

water:acetonitrile with 0.2% formic acid) was prepared by weighing 3.78 g of ammonium formate into a

1 L graduated cylinder. Water was added alongwith a stir bar and mixed to dissolve before diluting to

volume with water. The solutionwas transferred to a 2 L mobile phase container along with 1 L of

acetonitrile, 4mL of formic acid, stir bar, and then thoroughly mixed. Mobile phase B was also used for

the rinse solutions in the autosampler.

D.

Preparation of Standard Solutions

The carnitine stock standard was prepared at a concentration of 25mg/mL by weighing 0.25 g of L-

carnitine into a 20mL polypropylene tube followed by 10mL of water to dissolve. The purity of L-

carnitine from the Certificate of Analysis (CoA) and moisture determined by Karl Fisher immediately at