![Show Menu](styles/mobile-menu.png)
![Page Background](./../common/page-substrates/page0121.png)
any interference or carryover. The working standards, reagent blank, and reagent blank + internal
standard were included with each total analysis and treated the same as samples through the sample
preparation. Liners designated for the working standards received 50 µl of the appropriate intermediate
working standard level. All liners except the reagent blank received 50 µl of the intermediate internal
standard solution. A 5mL volume of water followed by 2.5mL of 70% (w/w) nitric acid delivered with a
bottle top dispenserwere then added to each liner, capped, and vortexed to mix. The microwave
program used was a ramp to temperature of 120°C over 10minutes, followed by a 40minute hold at a
power of 1000W, ending in a cool down (6). The contents of the vessels were transferred into 50mL
polypropylene tubes withwater and diluted to a volume of 25mL with water. A 0.5mL aliquot of the
sample solution was mixed with 0.5mL of acetonitrile in a microcentrifuge tube, and then filtered
through a 0.45 µm PTFE syringe filter into a silanized injection vial. Aliquots of 0.5mL of the working
standard and reagent blank solutions were mixed with 0.5mL of acetonitrile directly in the silanized
injection vials.
F.
LC/MS/MS Parameters
A Shimadzu Prominence liquid chromatography system equipped with an Agilent Zorbax 300-SCX
column (3.0 x 50mm, 5 µm) was utilized. A flow rate of 1.0mL/min was maintained over the 4.2minute
total run time. The mobile phase conditions were 100% mobile phase A until 1.0minute, ramped to
100% mobile phase B by 1.5minutes, and ramped back to 100% A by 3.0minutes. A column
temperature of 40°C, and an autosampler temperature of 5°C was maintained. A 1 µl injectionwas
used. Autosampler rinse settings were adjusted to eliminate carryover as much as possible. An ABSciex
API 4000mass spectrometer with positive ion electrospray (ESI) ionization was used in multiple reaction
monitoring (MRM) mode. The MS/MS overall settings used are described in Table
2015.10A
. The
MS/MS settings may need to be modified except for ionization, mode, and gas types to obtain optimum
chromatography and sensitivity. Figures
2015.10A
and
2015.10B
show typical extracted ion
chromatograms (XIC) of NIST SRM1849a for choline and carnitine.
G.
Quantification and Confirmation
The quantification of choline and carnitine was accomplished by generation of calibration curves using
the peak area ratio of the chosen transition (Table
2015.10B
) versus the corresponding deuterated
internal standards. Least square regression analysis using a linear model with 1/x
2
weighting was used
for both analytes. Confirmation was achieved through analysis of ion ratios between samples and
reference standards for at least one additional transition listed in the table. The concentration of each
analyte in a sample was calculated by the following equation:
= C × V × D ×
100 10
6
ng/mg
Candidates for 2016 Method of the Year
120