![Show Menu](styles/mobile-menu.png)
![Page Background](./../common/page-substrates/page0116.png)
1388
Bidlack et al.:
J
ournal of
AOAC I
nternational
V
ol.
98, N
o.
5, 2015
will be. Degas the mobile phase and postcolumn electrolyte
solutions by bubbling helium through them at a flow rate just
fast enough to cause small ripples on the surface of the mobile
phase and postcolumn solutions. To maximize the life of the
zinc reactor column, degas the mobile phase and postcolumn
electrolyte solution for at least 30 min before connecting the zinc
reactor column or do not pump mobile phase and postcolumn
electrolyte solutions until at least 30 min after degassing begins.
Once the mobile phase and postcolumn electrolyte solutions
have been degassed, allow the column and postcolumn reactor
to equilibrate with mobile phase flowing at 0.4 mL/min and
postcolumn electrolyte solution flowing at 0.4 mL/min for at
least 30 min prior to the first injection if the zinc reactor has
been used for previous analyses or several hours if the zinc
postcolumn reactor has been freshly packed. Once the mobile
phase and postcolumn solutions have been degassed, reduce the
helium flow rate so that only a small stream of helium bubbles
are visible in the mobile phase and postcolumn solutions and
there is minimal disturbance to the surface of these solutions.
Bubble helium very slowly through the mobile phase and
postcolumn electrolyte solutions continuously throughout the
entire run. Once the run has started, do not adjust the helium
flow rate. Allow the fluorescence detector lamp to warm up
30 min prior to the first injection. (
Note
: When the mobile phase
and postcolumn electrolyte solution are continuously sparged
with helium throughout a run, it is not necessary to pack the
postcolumn reactor with zinc at the beginning of every run. It
should be possible to analyze hundreds of samples before the
zinc reactor column must be repacked.)
(
3
)
HPLC of standards and samples
.—Inject the most
concentrated standard (approximately 80 µg/L) onto the
column and observe the response on the fluorescence detector.
If necessary, adjust the detector gain and sensitivity settings so
that the standard response is within the range of the detector.
Once the detector settings have been determined, inject the
most concentrated standard 3–4 times and note the peak areas.
If the system is equilibrated, the RSD of the standard peak areas
should be ≤2%, and the peak areas should not steadily increase
or decrease by more than 4% from the first injection to the third
or fourth injection. If the RSD is >2%, locate the source of the
imprecision and correct it before beginning the sample analysis.
If peak areas steadily increase or decrease by more than 4%, the
system is not equilibrated and must be allowed to equilibrate
longer. Once the system has reached equilibrium and the
RSD is ≤2%, inject a set of standards, unknown samples, and
another set of standards. Every set of unknown samples must be
bracketed by standards.
(
4
)
Instrument shut down
.—After analyzing a set of
samples, simultaneously turn off the flow on the mobile phase
and postcolumn electrolyte solution pumps. Remove the helium
sparge lines from the mobile phase and postcolumn electrolyte
solution and turn off the helium. Turn off the fluorescence
detector lamp.
F. Calculations
The vitamin K
1
concentrations of samples analyzed on the
HPLC system are determined by comparison of peak areas from
samples of known weight with the peak areas of standards of
known concentration. Because the
cis
and
trans
vitamin K
1
retention times may shift slightly during a run, peak areas must
be used to quantitate
trans
vitamin K
1.
(a)
Calculation of the standard concentrations:
(
) 1,000,000,000
(
)
1 2 3
1 2
3 4
C
W V V V
D D D D
s
= × × × ×
× × ×
where C
s
is the working standard concentration in µg/L; W is
the weight of standard in g; V
1
, V
2
, and V
3
are the aliquots of
stock, intermediate I, and intermediate standard II solutions,
respectively, in mL; 1,000,000,000 is the conversion factor
from g/mL to mcg/L; and D
1
, D
2
, D
3
, and D
4
are the dilution
volumes of the stock, intermediate I, intermediate II, and
working standard solutions, respectively, in mL.
(b)
Peak areas are measured with a data system. Before
calculating concentrations, review all chromatograms to make
sure that
cis
and
trans
vitamin K
1
are baseline separated and that
there are no interfering peaks.
Trans
vitamin K
1
concentrations
cannot be calculated for any samples with interfering peaks
or poor separation between the
cis
and
trans
isomers (
see
Figures
2015.09B–D
). Check the integration of the
cis
and
trans
vitamin K
1
peaks.
Cis
vitamin K
1
elutes 1 to 3 min before
trans
vitamin K
1
depending on the analytical column used. If the peak
Figure 2015.09D. Chromatogram of an adult nutritional.
Candidates for 2016 Method of the Year
115