OMB Meeting Book_9-11-14 - page 76

COMMENT PARAMETERS 
USER COMMENTS 
AOAC OMA 
2011.19 
Method Name/Manuscript Title 
Chromium, Selenium, and Molybdenum in Infant Formula and Adult 
Nutritionals by ICP MS 
Manuscript Reference 
In your experience using the method, does the 
method perform according to the method’s 
applicability as written? 
To compensate the signal enhancement of Se82  Methanol (Source of 
Carbon) is being added to Linearity Standards and Digested Sample.  
Methanol could interfere for Cr52  analysis (Ar40 + C12 = ArC52+ ).   
To achieve better sensitivity Se82   can be analyzed as SeO98  (Se82  + 
O16) in DRC mode using Oxygen as Reaction Gas. 
In your experience with the method, is there 
any safety concerns identified while using or 
regarding use of the method? 
None 
Do you have any information that supports 
regarding the reproducibility of the method as 
written:  If so, please specify and submit 
information? 
See point 1. 
Please attach any documentation to support 
your feedback, if available.   
Based on your experience with the method, 
are there any recommended changes to the 
AOAC First Action Method as written? 
See point 1.