Single Laboratory Validation – AOAC First Action Method 2012.22 Vitamin C in Infant Formula and Adult/Pediatric

Nutritional Formula by UPLC-UV

Date of Issue: February 14

th

, 2013

RDLS-RD130018

E.Campos Giménez (NRC/QS, Lausanne)

Page 6

Results

Single Laboratory Validation

System

suitability test

Relative standard deviation of retention time and response between the six

standard injections were below 1 % for all analytical series.

Specificity

All the placebo samples were free from chromatographic interferences

.

AU

0.000

0.002

0.004

0.006

0.008

0.010

0.012

0.014

0.016

0.018

0.020

Minutes

0.00 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40 2.60 2.80 3.00 3.20 3.40 3.60 3.80 4.00

AU

0.000

0.002

0.004

0.006

0.008

0.010

0.012

0.014

0.016

0.018

0.020

Minutes

0.00 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40 2.60 2.80 3.00 3.20 3.40 3.60 3.80 4.00

Child Formula Powder

Infant Elemental Powder

Ascorbic acid

Ascorbic acid

AU

0.000

0.002

0.004

0.006

0.008

0.010

0.012

0.014

0.016

0.018

0.020

Minutes

0.00 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40 2.60 2.80 3.00 3.20 3.40 3.60 3.80 4.00

AdultNutritional RTF High Protein

Ascorbic acid

AU

-0.002

0.000

0.002

0.004

0.006

0.008

0.010

0.012

0.014

0.016

0.018

0.020

Minutes

0.00 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40 2.60 2.80 3.00 3.20 3.40 3.60 3.80 4.00

Infant Formula RTF

Ascorbic acid

AU

0.000

0.002

0.004

0.006

0.008

0.010

0.012

0.014

0.016

0.018

0.020

Minutes

0.00 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40 2.60 2.80 3.00 3.20 3.40 3.60 3.80 4.00

AdultNutritional RTF High Fat

Ascorbic acid

Linearity/

Calibration Fit

Calibration curves were plotted and linearity demonstrated by R

2

> 0.9999 and

calibration errors well below 5 % for all levels used in daily operation (0.5-10

µg/mL) (Figure 2). This corresponds to 2.5–50 mg/100 g in reconstituted

samples. Although linearity up to 500 mg/100 g has been proven (data not

shown), we recommend diluting the samples if they contain more than 50

mg/100 g.

Figure 2: linearity range/calibration fit.

Continued on next page

7 / 20

VitC-03/2012.22