AOAC INTERNATIONAL

8. ADDITIONAL REAGENTS

109

Lysis reagent

– Using test kit reagents, pipette 150

µ

L of protease into one 12-mL bottle of lysis

110

buffer. Label bottle with the date prepared. Reagent will remain stable for up to two weeks if

111

stored at 2-8ºC.

112

113

9. APPARATUS

114

Incubators

– For maintaining media at 35

±

1ºC and 39-42ºC.

115

Stomacher

– Seward model 400 or equivalent for mixing the sponge sample with enrichment

116

media.

117

BAX

®

System apparatus

118

BAX® System Q7 cycler/detector with computer workstation

119

BAX® System application software

120

Cluster tubes with caps and racks for lysis

121

Capping/de-capping tools – for removing and sealing cluster tube caps and PCR

122

tube caps without jarring the contents

123

Heating and cooling blocks with inserts – for maintaining lysis tubes at 37±2ºC,

124

95±3ºC, and 2-8°C.

125

Pipettes – for transferring reagents; two adjustable mechanical pipettes covering 20-

126

200

µ

L and 5-50

µ

L; one repeating pipette; and one multi-channel pipette covering 8

127

channels and 5-50

µ

L. Pipettes should be calibrated to deliver required volumes

128

within 10%.

129

Pipette tips with barriers: 0.5-250

µ

L, 0.5-100

µ

L extended barrier; 5 mL repeater

130

pipette tips

131

Dupont BAX Salmonella PTM Report

Modification Approved 2012 / PTM Certification No. 081201

For Expert Review Panel Use Only