Table of Contents Table of Contents
Previous Page  247 / 1195 Next Page
Information
Show Menu
Previous Page 247 / 1195 Next Page
Page Background

AOAC INTERNATIONAL

EVALUATION OF THE BAX® SYSTEM REAL TIME ASSAY FOR THE

1

DETECTION OF

SALMONELLA

IN SELECTED FOODS

2

3

1. ABSTRACT

4

The BAX® System PCR assay for

Salmonella

was previously validated for the detection of this

5

pathogen in a variety of foods (AOAC- RI Performance Tested Method #100201, AOAC Official

6

Method of Analysis 2003.09, and Health Canada Compendium Method MFLP-29). A new BAX®

7

System test kit for

Salmonella

, with the proven track record of the previous test kit’s genetic target,

8

but the advantages of reformulation as a real-time assay, is proposed for validation in this study.

9

An evaluation was conducted on five food matrices and one environmental sample type to compare

10

the BAX® System real-time assay for

Salmonella

with the FDA-BAM, USDA-FSIS MLG, and Health

11

Canada Compendium culture methods for detecting

Salmonella

spp. Multiple reference methods

12

were used in this study to receive a validation applicable to the widest range of customers, who

13

require at least a minimal validation using their current reference method before performing internal

14

validations with a rapid method such as the BAX® System. All selected sample types (ground beef,

15

cream cheese, bagged lettuce, dry pet food, chicken carcass rinses, and stainless steel) were tested

16

by DuPont Qualicon, and one food and the environmental matrix were also tested by an independent

17

laboratory. No significant difference in test and reference method performance was found for any of

18

the matrices tested.

19

Inclusivity/exclusivity data was generated which demonstrated that the BAX® System real-time assay

20

is reactive for the major serotypes of

Salmonella

and non-reactive to other related non-

Salmonella

21

genera. Additionally, ruggedness data was generated to evaluate the performance of the assay to

22

deviations outside manufacturer’s recommendations, and accelerated stability studies and lot-to-lot

23

evaluations were performed which validated a rational test kit shelf life and demonstrated consistent

24

performance across multiple production lots of the assay.

25

26

27

28

Dupont BAX Salmonella PTM Report

PTM Certification No. 081201 (08/07/2012)

For Expert Review Panel Use Only

Do Not Distribute