AOAC INTERNATIONAL
(HE), Brilliant Green Sulfa (BGS) and Bismuth Sulfite (BS) agars and other appropriate
178
confirmatory media for reference method and alternate culture method confirmation.
179
11. Lysis reagent – Pipette 150
µ
L of protease into one 12-mL bottle of lysis buffer. Label bottle
180
with the date prepared. Reagent will remain stable for up to two weeks if stored at 2-8ºC.
181
182
14. SAFETY PRECAUTIONS
183
184
Kits
— Reagents used in the BAX® System
should pose no hazards when used as directed.
185
Dispose of lysate, PCR mixture and other waste according to your site practices.
186
Cycler/detector
— Only qualified laboratory personnel should operate the cycler/detector. Do not
187
attempt to repair the instrument. Live power may still be available inside the unit even when a
188
fuse has blown or been removed. Refer to the BAX® System User Guide for maintenance
189
procedures when cleaning the unit or changing a fuse. The heating block can become hot enough
190
during normal operation to cause burns or cause liquids to boil. Wear safety glasses or other eye
191
protection at all times during operation.
192
Enrichment Broths
— All enrichment broths may contain varying pathogens whether they contain
193
Salmonella
or not and thus should be sterilized and disposed of using proper procedures
194
following any culture-based confirmatory steps.
195
196
15. GENERAL PREPARATION / SAMPLE PREPARATION AND
197
RECOVERY
198
199
Note on Selection of Reference Methods
200
Since many United States end users require some degree of validation of a rapid method compared to
201
either the FDA-BAM or USDA-MLG as appropriate, and Canadian end users require the presentation
202
of similar data comparing the appropriate Compendium method to a proposed alternative method,
203
these studies were conducted using each of these three methods for at least one matrix comparison.
204
205
206
Dupont BAX Salmonella PTM Report
PTM Certification No. 081201 (08/07/2012)
For Expert Review Panel Use Only
Do Not Distribute