lysis step may be considered a potential biohazard and should NOT be inserted into the 3M

1

Molecular Detection Instrument.

2

3

After use, the enrichment medium and the 3M Molecular Detection Assay

Salmonella

tubes

4

can potentially contain pathogenic materials. When testing is complete, follow current

5

industry standards for the disposal of contaminated waste. Consult the Material Safety Data

6

Sheet for additional information and local regulations for disposal.

7

8

D.

MethodPreparation and Precautions

9

The 3M Molecular Detection Assay

Salmonella

is intended for use in a laboratory

10

environment by professionals trained in laboratory techniques. The user should read,

11

understand and follow all safety information in the instructions for the 3M Molecular

12

Detection System and the 3M Molecular Detection Assay

Salmonella

and retain the safety

13

instructions for future reference. Follow all instructions carefully. Failure to do so may lead

14

to inaccurate results.

15

16

Store the 3M Molecular Detection Assay

Salmonella

at 2-8°C. Do not freeze. Keep kit away

17

from light during storage. After opening the kit, check that the foil pouch is undamaged. If

18

the pouch is damaged, do not use. After opening, unused reagent tubes should always be

19

stored in the re-sealable pouch with the desiccant inside to maintain stability of the

20

lyophilized reagents. Store resealed pouches at 2-8°C for no longer than 1 month. Do not

21

use 3M Molecular Detection Assay

Salmonella

past the expiration date. Expiration date and

22

lot number are noted on the outside label of the box.

23

24

Use proper aseptic technique. Use proper precautions for biosafety level 2 microorganisms.

25

Periodically laboratory benches and equipment (pipettes, cap/decap tools, etc…) should be

26

decontaminated with a 1-5% (v:v in water) household bleach solution or DNA removal

27

solution.

28

29

30

E.

Sample Preparation

31

Table 1 presents guidance for the enrichment of food and feed samples at a 1:10 dilution. It

32

is the user’s responsibility to validate alternate sampling protocols or dilution ratios to

33

ensure this test method meets the user’s criteria.

34

1.

Pre-warm 3M BPW ISO enrichment medium to 37 ±1°C or 41.5 ±1°C.

35

2.

Aseptically combine the enrichment medium and sample. Homogenize thoroughly

36

for 2 minutes Incubate at 37 ±1°C or 41.5 ±1°C.

37

For all meat and highly particulate samples, the use of filter bags is recommended.

38

In an AOAC Research Institute Performance Tested Method study, the 3M Molecular Detection

39

Assay

Salmonella

was found to be an effective method for the detection of Salmonella in the

40

following matrices:

41