3M Molecular Detection Assay Salmonella Collaborative Study
July 2012
OMA-2012-July-XXX
4.6
Most Probable Number (MPN) Analysis.
To determine the level of
Salmonella
1
spp
.
in the inoculated food test portions on the day of initiation of analysis, a 5-
2
tube MPN determination is performed by the coordinating laboratory following
3
the appropriate reference method. From each of the inoculated food batches,
4
prepare 5 test portions of 100 g, 5 test portions of 25 g and 5 test portions of 10 g.
5
To each test portion, add the appropriate reference method enrichment media to
6
make a 1:10 dilution. Follow the reference method through to confirmations. Use
7
the number of positives from the high, medium and low levels to calculate the
8
MPN for each inoculated level of each matrix. Access the MPN calculator
9
( http://www.lcfltd.com/customer/LCFMPNCalculator.exe )to determine the MPN
10
values and 95% confidence intervals. Data should be entered from high level to
11
low. Report MPN as MPN/test portion.
12
13
5.0
3M Molecular Detection Assay
Salmonella
14
15
5.1
Applicability
16
17
For the detection of
Salmonella
species in food and feed matrices: raw ground
18
beef, cooked breaded chicken, raw shrimp, bagged spinach, wet pet food, and
19
pasteurized liquid whole egg.
20
21
Caution
:
The 3M Molecular Detection Assay
Salmonella
is intended for use
22
in a laboratory environment by professionals trained in laboratory techniques.
23
The user should read, understand and follow all safety information in the
24
instructions for the 3M Molecular Detection System and the 3M Molecular
25
Detection Assay
Salmonella
and retain the safety instructions for future
26
reference. Follow all instructions carefully. Failure to do so may lead to
27
inaccurate results.
28
29
The 3M Molecular Detection Instrument is intended for use with samples that
30
have undergone heat treatment during the assay lysis step, which is designed
31
to destroy organisms present in the sample. Samples that have not been
32
properly heat treated during the assay lysis step may be considered a potential
33
biohazard and should NOT be inserted into the 3M Molecular Detection
34
Instrument.
35
36
Store the 3M Molecular Detection Assay
Salmonella
at 2-8°C. Do not freeze.
37
Keep kit away from light during storage. After opening the kit, check that the
38
foil pouch is undamaged. If the pouch is damaged, do not use. After opening,
39
unused reagent tubes should always be stored in the re
‑
sealable pouch with
40
the desiccant inside to maintain stability of the lyophilized reagents. Store
41
resealed pouches at 2-8°C for no longer than 1 month. Do not use 3M
42
Molecular Detection Assay
Salmonella
past the expiration date. Expiration
43
date and lot number are noted on the outside label of the box.
44
45