3M Molecular Detection Assay Salmonella Collaborative Study
July 2012
OMA-2012-July-XXX
5.5.10
Refrigerator
. – capable of maintaining 2-8°C, for storing the 3M
1
Molecular Detection Assay
2
5.5.11
Computer
. – compatible with the 3M Molecular Detection Instrument
3
4
5.6
Determination
5
5.6.1
Enrichment
6
5.6.1.1
Raw ground beef
- aseptically add the 25g test portion to 225 mL
7
pre-warmed (37°C) 3M BPW ISO.
8
5.6.1.1.1
Homogenize thoroughly in a filter bag for 2 min.
9
5.6.1.1.2
Incubate 8 hr at 37 ± 1°C.
10
5.6.1.2
Wet dog food
- aseptically add the 375g test portion to 3375 mL
11
pre-warmed (37°C) 3M BPW ISO.
12
5.6.1.2.1
Homogenize thoroughly in a filter bag for 2 min.
13
5.6.1.2.2
Incubate 18 hr at 37 ± 1°C.
14
15
5.6.2
Prepare the 3M Molecular Detection speed loader tray, chill block insert,
16
heat block insert and instrument.
17
5.6.3
Transfer 20 µL enriched sample into a LS tube.
18
5.6.4
Transfer 20 µL Negative Control [NC] into a Lysis Solution [LS] tube
19
after all enriched samples have been completed.
20
5.6.5
Invert 3-5 times and heat 15±1 min at 100±1°C.
21
5.6.6
Place tubes (without rack lid) in chill block insert for 10±1 min.
22
5.6.7
Replace rack lid and invert 3-5 times. Allow to sit on bench at least 5 min.
23
5.6.8
Transfer 20 µL sample lysate from the upper portion of fluid in the LS
24
tube into regent tube. Mix gently by pipetting up and down 5 times.
25
5.6.9
Transfer 20 µL of NC lysate into a reagent tube.
26
5.6.10
Transfer 20 µL of NC lysate into a Reagent Control (RC) tube.
27
5.6.11
Load capped tubes into speed loader tray and close lid.
28
5.6.12
Start assay.
29
5.6.13
Results and Interpretation - An algorithm interprets the light output curve
30
resulting from the detection of the nucleic acid amplification. Results are
31
analyzed automatically by the software and are color-coded based on the
32
result. A Positive or Negative result is determined by analysis of a number
33
of unique curve parameters. Presumptive positive results are reported in
34
real-time while Negative and Inspect results will be displayed after the run
35
is completed. Presumptive positive results should be confirmed using your
36
preferred method or as specified by local regulations.
37
38
NOTE:
Even a negative sample will not give a zero reading as the
39
system and 3M Molecular Assay
Salmonella
amplification reagents
40
have a “background” relative light unit (RLU).
41
42
In the rare event of any unusual light output, the algorithm labels this
43
as “Inspect.” 3M recommends the user to repeat the assay for any
44
Inspect samples. If the result continues to be Inspect, proceed to
45
DRAFT DOCUMENT
9