3M Molecular Detection Assay Salmonella Collaborative Study

July 2012

OMA-2012-July-XXX

5.5.10

Refrigerator

. – capable of maintaining 2-8°C, for storing the 3M

1

Molecular Detection Assay

2

5.5.11

Computer

. – compatible with the 3M Molecular Detection Instrument

3

4

5.6

Determination

5

5.6.1

Enrichment

6

5.6.1.1

Raw ground beef

- aseptically add the 25g test portion to 225 mL

7

pre-warmed (37°C) 3M BPW ISO.

8

5.6.1.1.1

Homogenize thoroughly in a filter bag for 2 min.

9

5.6.1.1.2

Incubate 8 hr at 37 ± 1°C.

10

5.6.1.2

Wet dog food

- aseptically add the 375g test portion to 3375 mL

11

pre-warmed (37°C) 3M BPW ISO.

12

5.6.1.2.1

Homogenize thoroughly in a filter bag for 2 min.

13

5.6.1.2.2

Incubate 18 hr at 37 ± 1°C.

14

15

5.6.2

Prepare the 3M Molecular Detection speed loader tray, chill block insert,

16

heat block insert and instrument.

17

5.6.3

Transfer 20 µL enriched sample into a LS tube.

18

5.6.4

Transfer 20 µL Negative Control [NC] into a Lysis Solution [LS] tube

19

after all enriched samples have been completed.

20

5.6.5

Invert 3-5 times and heat 15±1 min at 100±1°C.

21

5.6.6

Place tubes (without rack lid) in chill block insert for 10±1 min.

22

5.6.7

Replace rack lid and invert 3-5 times. Allow to sit on bench at least 5 min.

23

5.6.8

Transfer 20 µL sample lysate from the upper portion of fluid in the LS

24

tube into regent tube. Mix gently by pipetting up and down 5 times.

25

5.6.9

Transfer 20 µL of NC lysate into a reagent tube.

26

5.6.10

Transfer 20 µL of NC lysate into a Reagent Control (RC) tube.

27

5.6.11

Load capped tubes into speed loader tray and close lid.

28

5.6.12

Start assay.

29

5.6.13

Results and Interpretation - An algorithm interprets the light output curve

30

resulting from the detection of the nucleic acid amplification. Results are

31

analyzed automatically by the software and are color-coded based on the

32

result. A Positive or Negative result is determined by analysis of a number

33

of unique curve parameters. Presumptive positive results are reported in

34

real-time while Negative and Inspect results will be displayed after the run

35

is completed. Presumptive positive results should be confirmed using your

36

preferred method or as specified by local regulations.

37

38

NOTE:

Even a negative sample will not give a zero reading as the

39

system and 3M Molecular Assay

Salmonella

amplification reagents

40

have a “background” relative light unit (RLU).

41

42

In the rare event of any unusual light output, the algorithm labels this

43

as “Inspect.” 3M recommends the user to repeat the assay for any

44

Inspect samples. If the result continues to be Inspect, proceed to

45

DRAFT DOCUMENT