Silliker
Chem, Res. Center Crete, IL – Report of A Validation of LC-MS/MS Method for Folate Analysis
31
Table 22: 5-formyl-THF analysis in BCR 485 (dried vegetable powder) over a period of about 3
months
5-CHO-THF
Results
09/11/12
09/14/12
09/18/12
10/19/12
10/26/12
11/02/12
mcg/g
Mean
0.063
0.070
0.073
0.065
0.059
0.071
%RSD
10.6%
51.1%
12.1%
2.9%
18.7%
12.2%
N =
3
3
3
3
3
3
5-CHO-THF
Results
11/07/12
11/13/12
12/27/12
Grand
Grand SD
mcg/g
Mean
0.075
0.052
0.054
0.063
0.010
%RSD
5.2%
13.0%
3.2%
15.6%
N =
3
3
6
25
25
There is no value provided for 5-formyl-THF for this sample by BCR
Total folate as folic acid results obtained for the BCR 485 (dried vegetable powder) by the method
over a period of about 3 months are presented in Table 29. The results obtained were quite
consistent over period of analysis. Values obtained for total folate as folic acid was about 96% of
the lowest range specified for this reference material. Interday as well as intraday precision of the
method over a period of about 3 months was found to be satisfactory.
Table 23: Total folate as folic analysis in BCR 485 (dried vegetable powder) over a period of
about 3 months
Vitamers sum as Folic
09/11/12 09/14/12 09/18/12 09/21/12 10/10/12 10/12/12 10/19/12
mcg/g
Mean
2.312
2.465
2.208
2.494
3.018
3.333
2.931
%RSD
3.5%
2.3%
1.0%
3.5%
2.2%
8.0%
1.9%
N =
3
3
3
3
3
3
3
Vitamers sum as Folic
10/26/12 11/02/12 11/07/12 11/13/12 12/27/12 Grand
Grand
SD
mcg/g
Mean
2.922
3.195
2.654
2.422
2.716
2.710
0.339
%RSD
0.1%
2.3%
2.4%
0.6%
3.0%
12.5%
N =
3
2
3
3
6
38
38
Certified Total Folate
Value*
3.15 + 0.28
*BCR Certified Total Folate value as folic acid obtained by microbiological assay (L. rhamnosus)
Method Accuracy Evaluation Repeated:
The method accuracy was repeated again recently (API 5500 from ABSciex). The results are presented
in the following Table. The results of this investigation again help in establishing the accuracy of the
method.