Table of Contents Table of Contents
Previous Page  805 / 1096 Next Page
Information
Show Menu
Previous Page 805 / 1096 Next Page
Page Background

S789

ESTRO 36

_______________________________________________________________________________________________

considering all the techniques, only 4.4% of mean g

mean

tests resulted out of tolerance. In addition, removing

class-2 errors, this percentage decreases to approximately

3%. Actually the workload of IVD procedures on 9 patients

is 1 hour per day.

Conclusion

IVD performed using SOFTDISO assures: (i) a rapid response

of dose delivery alert with a reduced workload; (ii) a large

number of patients tested daily and (iii) for out- of-

tolerance tests repeating IVD in the subsequent day, the

possibility to verify the efficacy of the adopted

corrections.

EP-1477 Evaluating gamma-index quality assurance

methods for Nasopharynx Volumetric Arc Therapy

(VMAT)

E.M. Pogson

1,2,3

, S. Arumugam

2

, S. Blake

1

, N. Roberts

4

, C.

Hansen

5,6

, M. Currie

7

, M. Carolan

7

, P. Vial

2

, J. Juresic

2

, C.

Ochoa

2

, J. Yakobi

2

, A. Haman

2

, A. Trtovac

2

, L.

Holloway

1,2,3,4,8

, D.I. Thwaites

1

1

University of Sydney, Institute of Medical Physics-

School of Physics- Faculty of Science, Sydney NSW,

Australia

2

South Western Sydney Local Health District, Liverpool

and Macarthur Cancer Therapy Centres, Liverpool,

Australia

3

Ingham Institute, Medical Physics, Liverpool, Australia

4

University of Wollongong, Centre for Medical Radiation

Physics- School of Physics, Wollongong, Australia

5

Odense University Hospital, Laboratory of Radiation

Physics, Odense, Denmark

6

University of Southern Denmark, Faculty of Health

Sciences- University of Southern Denmark- Denmark,

Odense, Denmark

7

Illawarra and Shoalhaven Local Health District,

Illawarra Cancer Care Centre, Wollongong, Australia

8

University of New South Wales, South Western Sydney

Clinical School, Sydney, Australia

Purpose or Objective

Pre-treatment dose verification is often performed on

dose measuring phantoms with some form of gamma

evaluation. However it has been shown that the clinical

relevance of a 3% and 3mm pass rate tolerance is

questionable. The purpose of this study is to simulate

machine errors of clinical significance for nasopharynx

patients and test if these errors can be detected on a

standard commercial phantom. In this study systematic

errors including collimator rotation, gantry rotation, MLC

shifts, and MLC field sizes are investigated.

Material and Methods

Ten retrospective VMAT patients were planned with a

department protocol. Machine errors were deliberately

introduced to all plans. Plans were modified by increments

using Python to create simulated error plans; -5 to 5° for

gantry and collimator angles and -5 to +5mm for MLC shift

and MLC field size, considering each parameter

separately. Simulated error plans (Dose

error

) were

compared to the original non-error plan (Dose

Baseline

)

utilising

equation

(1).

(1)

All error plans doses were then recalculated in Pinnacle

3

.

Plans were reviewed against acceptable tolerance limits.

Plans were above tolerance and considered unacceptable

if PTV D95%, Brainstem D1cc or spinal cord D1cc were

beyond a ±5% deviation in dose. Additionally if either of

the left or right parotid mean doses were beyond ±10%,

this was also considered an unacceptable plan.

The smallest unacceptable error plan for each error type

(including the Gantry (G), Collimator (C), MLC Shift (S),

and MLC Field Size (F) error was delivered on an Elekta

Linac and dose was measured using an ArcCheck. Gamma

analysis was performed in SNCpatient version 6.6 utilising

a global 3%/3 mm (10% threshold with correction off)

gamma pass rate. Before measurement, the Linacs were

tested for MLC, gantry and dose accuracy. Only one

patient’s -5° gantry error was deemed unacceptable and

subsequently measured, patient 2 with this error detected

(gamma pass rate of 68.8%).

Results

The results for 10 patients are shown in Figure 1.

Figure 1. Gamma pass rate (%) for non-error (NE) plans

and for deliberately introduced errors (including the

Collimator (C), MLC shift (S), and MLC Field Size (F)

error), where the latter are selected as exceeding dose

tolerances by the smallest magnitude. *Patient 1 F error

of -5 was not deliverable due to machine tolerances,

hence an F error of -2 is utilised here.

The global 3%/3mm gamma pass is able to detect the

majority of unacceptable plans, however some MLC field

size plans still pass. Decreasing the MLC field size by 1mm

can result in significantly reduced dose to the PTV, which

affects tumour control e.g. patient 3 MLC FS-1 passed,

however this plan under-doses the PTV63Gy by -5.4%

relative to the original non-error plan.

Conclusion

Not all deliberately introduced clinically significant errors

were discovered for VMAT plans using a typical 3%/3mm

(10% threshold with correction off) gamma pass rate.

EP-1478 A split field beam model of Beam Modulator

linear accelerator in Pinnacle treatment planning

system

M. Chandrasekaran

1

, S. Worrall

1

, M.K.H. Chan

1

, N.

Khater

1

, C. Birch

1

1

University Hospital Southampton NHS Foundation Trust,

Radiotherapy Physics, Southampton, United Kingdom