Table of Contents Table of Contents
Previous Page  346 / 453 Next Page
Information
Show Menu
Previous Page 346 / 453 Next Page
Page Background

mericon

E. coli Detection Workflows 8/2015

28

Table 9. Cycling protocol for Rotor-Gene Q

Step

Time

Temperatur

e

Comment

Initial PCR activation

step

5 min

95°C

Activation of HotStarTaq

Plus

DNA Polymerase

3-step cycling

Data collection at 60°C

for green, crimson,

orange and yellow

channels

Denaturation

15 s

95°C

Annealing

*

15 s

60°C

*

Extension

10 s

72°C

Number of cycles

40

*

Gain optimization before first acquisition at 60°C for green, crimson,

orange and yellow channels

8.

Click “OK” to close the window and return to the Wizard.

9.

To set the gain optimization settings for the Green, Yellow, and Orange

Channels click “Gain Optimisation”.

10.

Select the 3 channels in the drop-down menu and click “Add”.

11.

In the dialog box that opens, confirm the standard settings. Click

“Perform Optimisation Before 1st Acquisition”. Then close the window.

Make sure that the tube at position 1 is not empty, since the gain

optimization will be performed on this tube.

OMAMAN-36 C: Method User Guide/Package Insert

ERP Use Only

January 2017

AOAC Research Institute

Expert Review Panel Use Only