Table 2014.04E. Mass spectrometer (Xevo TQ-S) operating conditions

Ionization mode

ESI positive

Capillary voltage

2.0 kV

Collision gas pressure

2–5 × 10

-3

mtorr

Source temperature

150°C

Source offset

30 V

Desolvation temperature 550°C

Cone gas flow

300 L/h

Desolvation flow rate

1000 L/h

Peak width at half height

0.7 amu

a

a

Instrumental resolution parameters for the TQ-S are set up by IntelliStart to

achieve a resolution of approximately 0.7 amu across the mass range.

Table 2014.04F. Chromatographic parameters

Mobile phase A

10 mM ammonium acetate [water-acetonitrile (95 + 5, v/v)]

Mobile phase B

10 mM ammonium acetate [water-acetonitrile (5 + 95, v/v)]

Flow rate

0.7 mL/min (analytical)

Flow rate into MS

Full flow

Column

Acquity UPLC BEH HILIC 1.7 µm, 2.1 × 100 mm

Column temperature 25°C

Injection volume

10 µL

Injection type

Full loop

Sample temperature 8°C

Candidates for 2016 Method of the Year