Table of Contents Table of Contents
Previous Page  215 / 220 Next Page
Information
Show Menu
Previous Page 215 / 220 Next Page
Page Background

Single Lab Validation of a Method for the Determination of β‐

Galactooligosaccharides in Infant Formula & Adult Nutritionals / Sean Austin (NRC, 

Lausanne)  

25 Jul 2016

CONFIDENTIAL

 ∙ This document may not be reproduced or disclosed to third parties without prior authorization 

Page 32 / 35

4.

PROTOCOL FOR MW DETERMINATION OF GOS BY UHPLC‐FLD‐MS 

Sample preparation: 

The same sample preparation is the same for both quantitative analysis and for the MW assignment 

of signals, thus follow the normal sample preparation procedure (the same vial prepared for the 

quantitative determination of GOS can thus also be used for the analysis of MW assignment).  

Alternatively one may prepare a sample using only the GOS ingredient.  In case the mass 

spectrometer has insufficient sensitivity it is possible to prepare a sample having ten times greater 

GOS concentration for the purposes of peak identification only (in this case it is recommended to 

use the GOS ingredient). 

Mass Spectrometer Set‐up: 

In addition to the UHPLC‐FLD instrument, an API 4000 QTRAP (Applied Biosystems) mass 

spectrometer detector was attached via a Triversa Nanomate (Advion) to work in nanoelectrospray 

mode.  The UHPLC flow was split at the exit of the LC column, approximately half of the flow (0.3 

mL/min) goes to the FLD, and the remaining 0.3 mL/min is directed to the nanomate source).  The 

nanomate further splits the flow down to about 600 nL/min for introduction to the mass 

spectrometer. The following describes the setup of the MS system in our lab.  The mass 

spectrometer settings in other labs should be optimised locally.   

NOTE: Do not connect the FLD and the mass spectrometer in series. In general the FLD cells cannot withstand the 

generated backpressure 

LC parameters: 

Use the same LC conditions as described in the quantitative method 

Triversa Nanomate source parameters: 

Method type:  LC Chip Coupling 

Trigger acquisition when input singnal received: Yes.   Signal: DigIn 1 

Voltage: 1.40 kV 

Polarity: Negative 

Output contact closure signal: MS Relay  

Duration: 2.5s 

Always use new nozzle:No 

VALIDATION REPORT

FOR ERP USE ONLY

DO NOT DISTRIBUTE