Validation of a LC-MS/MS method of Analysis of Nicotinic acid and Nicotinamide in infant formula and adult

nutritional samples

___________________________________________________________

Copyright © 2012 Silliker, Inc.

2

Principle

Free as well as bound nicotinic acid and nicotinamide are extracted using a slightly

acidic buffer employing protease and takadiastase treatments. The sample extracts are

analyzed using liquid chromatography (LC) with tandem mass spectrometric detection.

Reagents

a)

Deionized (DI) water.—Type I reagent grade, 18 MΩ-cm resistance or better.

b)

Ammonium Hydroxide.— Fisher Scientific; P/N A669-500, or equivalent

c)

Extraction Buffer - 12 mM Ammonium formate in water; pH = 5.0

d)

Ammonium Formate – Optima LC/MS Fisher Scientific A115-50g

e)

Formic Acid – Optima LC/MS Fisher Scientific A117-50mL

f)

Acetonitrile – Optima LCMS grade A955-4

g)

Protease – Sigma P3910

h)

Takadiastase – Pfaltz & Bauer P/N:T00040

Standards

a)

Nicotinic acid (B3-acid) -- 99% reagent grade or better (USP P/N #1461003)

b)

Nicotinamide (B3-amide) -- 99% reagent grade or better (USP P/N #1462006)

Internal Standards

a)

Internal Standard (IS) Nicotinic Acid-[13C3, 15N] – IsoSciences P/N 13257; 1mg

b)

Internal Standard (IS) Nicotinamide-[13C3, 15N] – IsoSciences P/N 14327; 1mg

Instrumentation

a)

Chromatographic system.— ABSciex 5500 QTrap / ABSciex 6500 QTrap

b)

Dionex Ultimate 3000 RS Pump, Autosampler, Column Compartment &Degasser

Thermo Fisher / Agilent 1290 Infinity Pump, Autosampler, column compartment &

Degasser

c)

Agilent 1290 Infinity Pump, Autosample, column compartment & Degasser

d)

Software.—AbSciex Analyst and MultiQuant chromatography data system

e)

Analytical column.— Thermo Acclaim C30 3um 2.1 x 150mm; (P/N 075725)

Labwares

a)

Autosampler vials.—Agilent 2.0ml amber vials, Fisher P/N 50-929-364)

VALIDATION REPORT