Table of Contents Table of Contents
Previous Page  796 / 1082 Next Page
Information
Show Menu
Previous Page 796 / 1082 Next Page
Page Background

S780

ESTRO 36 2017

_______________________________________________________________________________________________

patient’s -5° gantry error was deemed unacceptable and

subsequently measured, patient 2 with this error detected

(gamma pass rate of 68.8%).

Results

The results for 10 patients are shown in Figure 1.

Figure 1. Gamma pass rate (%) for non-error (NE) plans

and for deliberately introduced errors (including the

Collimator (C), MLC shift (S), and MLC Field Size (F)

error), where the latter are selected as exceeding dose

tolerances by the smallest magnitude. *Patient 1 F error

of -5 was not deliverable due to machine tolerances,

hence an F error of -2 is utilised here.

The global 3%/3mm gamma pass is able to detect the

majority of unacceptable plans, however some MLC field

size plans still pass. Decreasing the MLC field size by 1mm

can result in significantly reduced dose to the PTV, which

affects tumour control e.g. patient 3 MLC FS-1 passed,

however this plan under-doses the PTV63Gy by -5.4%

relative to the original non-error plan.

Conclusion

Not all deliberately introduced clinically significant errors

were discovered for VMAT plans using a typical 3%/3mm

(10% threshold with correction off) gamma pass rate.

EP-1478 A split field beam model of Beam Modulator

linear accelerator in Pinnacle treatment planning

system

M. Chandrasekaran

1

, S. Worrall

1

, M.K.H. Chan

1

, N.

Khater

1

, C. Birch

1

1

University Hospital Southampton NHS Foundation Trust,

Radiotherapy Physics, Southampton, United Kingdom

Purpose or Objective

The aim of this study was to model a beam modulator

linear accelerator in Pinnacle v.14.0 treatment planning

system for intracranial stereotactic radiosurgery and

radiotherapy.

Material and Methods

Depth dose, beam profile and total scatter correction

factor data were collected for 6 MV photons of Elekta

Synergy Beam Modulator

TM

linear accelerator with 80

leaves each of 4 mm leaf pitch using unshielded IBA

stereotactic field diode for field sizes ranging from 0.8x0.8

cm

2

to 4x4 cm

2

and field sizes above 4x4 cm

2

up to 16x21

cm

2

using IBA CC04 pinpoint chamber. The measured data

were imported in to the photon physics module of Pinnacle

v.14.0 and physical accelerator head specific data such as

primary collimator, flattening filter, MLC were input in

addition to beam data measurements. The auto modeler

of Pinnacle TPS was iteratively used to adjust parameters

such as photon beam energy spectrum, Gaussian height

and width of the photon source that affect various regions

of the depth doses and beam profiles to match measured

data. Dose grids of 1 mm and 2.5 mm were used for beam

modelling of fields from 0.8x0.8 cm

2

up to an equivalent

square field of 6.7 cm

2

and above 6.7 cm

2

respectively. A

common photon energy spectrum did not prove sufficient

to achieve the required agreement between Pinnacle

calculated and measured depth doses and beam profiles

for the whole range of field sizes. This was overcome by a

split field model that employs field size specific beam

energy spectra, with higher relative weights of low energy

bins and lower relative weights of high energy bins for

small fields and vice-versa for field sizes larger than 6.7

cm

2

. The validity of the model was tested independently

using a Standard Imaging Exradin A26 chamber in LUCY

phantom for field sizes ranging from 0.8x0.8 cm

2

by

comparing calculated and measured absolute doses and

relative output factors.

Results

Optimization of photon beam energy spectrum specific to

small field sizes improved the agreement of depth doses

both in and beyond build-up region for the small fields.

Measured versus calculated absolute planned doses were

found to be within 1% for field sizes larger than 1.6x1.6

cm

2

and less than 2.5% for 0.8x0.8 cm

2

field. The

agreement between the measured and calculated relative

output factors were within 2% for field sizes larger than

1.6x1.6 cm

2

and less than 3.5% for 0.8x0.8 cm

2

fields.