Table of Contents Table of Contents
Previous Page  67 / 154 Next Page
Information
Show Menu
Previous Page 67 / 154 Next Page
Page Background

3.  Is there information 

demonstrating that the method 

performs within the SMPR Method 

Preformance Requiements table 

specifications for all analytes in the 

SMPR applicability statement?  If 

not, please specify what is missing 

and whether or not the method's 

applicaiblity should be modified.  

For linearity ‐ yes.

For analytical range ‐ mostly, but limited to 25%, but does not meet the range 

requirements of greater than 50% (there are products out there that high).

For LOQ ‐  the SMPR specifies that the LOQ should be less than 0.1% in the sample 

materials. The LOQ specified in this method is ug/mL. using the lowest dilution for 

this method the curcumin (which is the highest LOQ ‐ 2.86 ug/mL) would back 

calculate to 0.0143 %, assuming that the sample mass is 500 mg, which is not 

actually specified in the method (taken from optimization information). So Yes, LOQ 

is sufficient.

Accuracy/Recovery ‐ Not specified in the document.

Repeatability ‐ As specified above the methodology used biases the repeatability 

data due to between injection and between sample being treated as replicates. 

Additionally, the CurcuViva product is consistently not within the <5% RSDr as 

specified in the SMPR. All other products appear to meet SMPR repeatability 

requirements, but I would recommend re‐treating the data and only taking the 

triplicate measurements to ensure true repeatability is obtained.

Some of the other dietary ingredients listed in Table 2 have also not been used in 

any of the multi‐component products.

1.  Based on the supporting 

information, were there any 

additional steps in the evaluation 

of the method that indicated the 

need for any addional 

precautionary statements in the 

method?

No

2.  Does the method contain 

system suitability tests or controls 

as specified by the SMPR?  If not, 

please indicate if there is a need 

for such tests or controls, and 

which ones.

The system suitability requirements of the SMPR specifies that check standards 

should be run throughout the validation ‐ in the case of this run, there were only 

blank samples, but no QC/check standards run. System suitability was not performed 

at the beginning of the validation either to ensure that the system was working 

properly.

3.  Is there information 

demonstrating that the method 

system suitability tests and 

controls as specified in the SMPR 

worked appropriately and as 

expected?  If no, please specify.

Linearity was sufficient, which would ensure that the system was suitable, but there 

was no actual system suitability performed.

4.  Based on the supporting 

information, is the method written 

clearly and concisely?  If no, please 

specify the needed revisions.

There is some information that is missing in the method preparation. The actual 

sample masses of the samples (or atleast recommended ranges of masses) are 

missing. This would cause a lot of confusion adopting the method. Additionally, 

several different extraction volumes were used. It would make sense to specify 

which volumes were suitable for what types of samples (or what expected ranges or 

ingredients). As for softgels, the results obtained from this would likely be 

mg/capsule, rather than mg/g, as you should not typically include the mass of the 

softgel in the analysis.

5.  Based on the supporting 

information, what are the 

pros/strenghts of the method?

The strengths:

simple sample preparation

fast chromatographic separation

baseline separation of the curcuminoids

suitable for many sample types, ingredients, etc.

6.  Based on the supporting 

information, what are the cons 

/weaknesses of the method?

The weaknesses:

lack of sufficient information for sample preparation

requires a significant amount of glassware

extraction solvent selection (see below)

IV.  General Submission Package