7 . Any general comments about 

the method?

The selection of extraction solvent seemed to be based more on the precision of the 

replicates rather than on analyte recovery. When looking at the error, the error of 

the methanol is slightly larger, but there is a significant difference between the 

quantity of curcumin and DMC extracted with methanol compared with the 80% 

methanol. Without any recovery data in this study, it appears that the accuracy of 

this method may be insufficient. There were also re‐extraction repetitive dilution 

study, but without descriptions of the actual methodologies, it is difficult to 

understand how this graph explains that only 1 extraction is required.

Do you recommend this method be 

adopted as a First Action and 

published in the Official Methods 

of Analysis of AOAC 

INTERNATIONAL?  Please specify 

rationale.  

No. The method appears to be missing key information including method 

recovery/accuracy. The repeatability data was calculated using duplicate injections 

as replicates and the selection of extraction solvent appears to impact the recovery 

of the curcuminoids.

V.  Recommendation for the Method