Table of Contents Table of Contents
Previous Page  422 / 1096 Next Page
Information
Show Menu
Previous Page 422 / 1096 Next Page
Page Background

S407

ESTRO 36

_______________________________________________________________________________________________

Conclusion

We have obtained an EPR signal for our solid polymer

dosimeter. The EPR signal increases linearly with dose for

the medical dose range, but it saturates for higher doses.

Although it is not comparable to the EPR dosimetry using

alanine, this signal could be a source of improved

understanding of the underlying dosimetric characteristics

of this material and it may be a supporting feature to the

optical signals from the dosimeter. We further foresee

interesting applications in particle therapy beams since

the signal production in solid-state dosimeters are

generally dependent on the ionization density.

PO-0769 A microDiamond for determination of

absorbed dose around high-dose-rate 192Ir

brachytherapy sources

V. Kaveckyte

1

, A. Malusek

1

, H. Benmaklouf

2

, G. Alm

Carlsson

1

, A. Carlsson Tedgren

2

1

Linköping University, Radiation Physics IMH, Linköping,

Sweden

2

Karolinska University Hospital, Radiation physics,

Stockholm, Sweden

Purpose or Objective

Experimental dosimetry of high-dose-rate (HDR)

192

Ir

brachytherapy (BT) sources is complicated due to steep

dose and dose-rate gradients, high dose rates and

softening of photon energy spectrum with depth. A single

crystal synthetic diamond detector PTW 60019 (marketed

as microDiamond) (PTW, Freiburg, Germany) has a small

active volume and was designed for such measurements in

high energy photon, electron and proton beams. It can be

read out directly with standard electrometers used at

radiotherapy departments, unlike thermoluminescent

detectors, which are currently the most used dosimeters

in BT but have to be pre- and post-processed with

dedicated equipment. Hence the purpose of this study was

to evaluate the suitability of a microDiamond for the

determination of absorbed dose to water in an HDR

192

Ir

beam quality. The use of three microDiamond samples also

allowed for assessment of their individual reproducibility.

Material and Methods

In-phantom measurements were performed using the

microSelectron HDR

192

Ir BT treatment unit. Oncentra

treatment planning system (TPS) was used to create

irradiation plans for a cubical PMMA phantom with a

microDiamond positioned at one of the three source-to-

detector distances (SDDs) (1.5, 2.5 and 5.5 cm). The

source was stepped by 0.5 cm over the total length of 6

cm to yield absorbed dose of 2 Gy at the reference point

of the detector. A phantom correction factor was applied

to account for the difference between the experimental

phantom and the spherical water phantom used for

absorbed dose calculations made with the TPS. The same

measurements were repeated for all three detectors

(mD1, mD2, mD3).

Results

Experimentally determined absorbed dose to water

deviated from that calculated with the TPS from -1 to +2

% and agreed to within experimental uncertainties for all

the detectors and SDDs (Figure 1). The mD2 overestimated

absorbed dose to water by up to 2% compared with the

estimates by the other two detectors. A decrease in the

difference with increasing SDD suggests that it might be

related to differences in the position of the active volume

inside the detector which is of higher importance closer to

the source where dose gradients are steeper. The

combined relative uncertainty in experimentally

determined absorbed dose to water did not exceed 2% ()

for all the detectors and SDDs. A variation in raw readings

was within 2% over the investigated range.

Conclusion

Preliminary results indicate that the dosimetric properties

of a microDiamond obviate the need for multiple

correction factors and facilitate dosimetry of HDR

192

Ir BT

sources. This, together with the convenience of use, shows

high potential of a microDiamond for quality assurance of

HDR BT treatment units at clinical sites. It must be noted,

nevertheless, that individual characterization of a

microDiamond is required to achieve high accuracy.

PO-0770 The distortions of the dose response

functions of dosimeters in the presence of a magnetic

field

H.K. Looe

1

, B. Delfs

1

, D. Harder

2

, B. Poppe

1

1

Carl von Ossietky University, University Clinic for

Medical Radiation Physics, Oldenburg, Germany

2

Georg August University, Prof em.- Medical Physics and

Biophysics, Göttingen, Germany

Purpose or Objective

The new developments of MRgRT have opened new

possibilities for high precision image-guided radiotherapy.

However, the secondary electrons liberated within the

medium by the primary photon beam are subjected to the

Lorentz force. Therefore, the trajectories of the

secondary electrons in non-water media, such as an air-

filled cavity or a high-density semiconductor, will differ

significantly from that in water. In this work we

demonstrate, using simple geometries, that the shape of

the lateral dose response functions of clinical detectors

will depend on the electron density of the detector

material, the beam quality and the magnetic field. The

dosimetric implications are discussed and correction

strategies are proposed.

Material and Methods

Based on the convolution model (Looe et al 2015), the one-

dimensional lateral dose response function, K(x-ξ), acting

as the convolution kernel transforming the true dose

profile D(ξ) into the measured signal profile M(x), was

derived by Monte-Carlo simulation for a simple cylindrical

detector placed at 5 cm depth in water using

60

Co and 6

MV slit beams. The cylinder with 1.13 mm radius and 2 mm

height was filled with water of normal density (1 g/cm

3

),

low density (0.0012 g/cm

3

) and enhanced density (3

g/cm

3

), where the latter two represent the density of an

air-filled ionization chamber and a semiconductor

detector respectively. Simulations were performed using