Table of Contents Table of Contents
Previous Page  425 / 1096 Next Page
Information
Show Menu
Previous Page 425 / 1096 Next Page
Page Background

S410

ESTRO 36

_______________________________________________________________________________________________

shapes and simulate organ deformations during RT. In this

abstract we propose a new, reusable 3D dosimetry system

based on OSL material embedded homogenously inside a

transparent matrix.

Material and Methods

Cuvette-sized prototypes of the dosimeter were produced,

consisting of a matrix; 4 g of a transparent silicone

elastomer (SE) (Sylgard 184, Dow Corning), and a

homogeneously embedded OSL material; 0.3 g of lithium

fluoride (LiF) doped with magnesium, copper and

phosphorus (LiF:Mg,Cu,P - MCP).

Three samples were prepared in standard OSL-reader

aluminum trays; a reference sample with silicone

elastomer, and two samples with OSL powder embedded

in the SE matrix, containing 0.06 mg and 0.2 mg OSL

powder (sample 1 and 2 respectively). They were read-out

using a Risø TL/OSL DA-20 reader. Samples were irradiated

with 1 Gy beta radiation and stimulated for 100 s with blue

light emitting diodes (LEDs), with emission centered at 470

nm and an intensity of ~80 mW/cm

2

.

Results

The transparency of the dosimeter (see Fig. 1) depended

on the concentration of MCP powder, which must be

optimized as a compromise between signal level per

volume and overall transparency. The refractive-index

match between LiF and the SE is quite good for visible

wavelengths, which minimizes light scattering from the

particles.

Approximately 10,000 and 40,000 counts were detected in

1 second per 1mm

3

voxel from samples 1 and 2,

respectively, corresponding to the anticipated signal

levels. Also, the silicone matrix in itself did not add to the

OSL signal (see Fig. 2). 3D distributions can be obtained

without the need for inversion algorithms, for example, by

stimulating the OSL dosimeter with a light sheet (from a

laser source), and imaging the luminescence intensity

across that sheet (by a combination of optical filters and

a camera), and shifting this plane across the dosimeter.

Conclusion

A new 3D dosimeter system based on OSL material has

been presented. It has the potential to verify complex 3D

RT doses with high spatial resolution, while maintaining

the advantages known from personal-dosimetry use of

OSL.

PO-0774 Investigation of dose-rate dependence at an

extensive range for PRESAGE radiochromic dosimeter

E.P. Pappas

1

, E. Zoros

1

, K. Zourari

2

, C.I. Hourdakis

2

, P.

Papagiannis

1

, P. Karaiskos

1

, E. Pantelis

1

1

National and Kapodistrian University of Athens, Medical

Physics Laboratory - Medical School, Athens, Greece

2

Greek Atomic Energy Commission, Division of Licensing

and Inspections, Athens, Greece

Purpose or Objective

The purpose is to investigate dose-rate dependence

effects for a recent formulation of the commercially

available PRESAGE radiochromic dosimeter (Heuris Inc,

NJ) in a wide range of dose delivery rates extending to

three orders of magnitude (0.018 – 19 Gy/min).

Material and Methods

In order to achieve an extensive dose rate range, this work

was divided into two separate studies. Lower dose rates

were delivered by

60

Co beams while higher dose rates were

achieved by a flattening-filter-free (FFF) linear

accelerator. For the low dose rate part of this study, 10

PMMA cuvettes (1×1×4 cm

3

), filled with PRESAGE samples,

were irradiated to the same dose with 5 different dose

rates. Irradiations were performed with a

60

Co PICKER unit

in a secondary standard calibration laboratory. The

samples were divided into groups of two and each group

was placed at a different distance (56.65 - 427 cm) from

the

60

Co source at a 5cm depth within a water phantom.

Irradiation times varied in order to deliver the same dose

of 1 Gy at the center of all cuvettes with dose rates in the

range of 0.018 – 1.0 Gy/min. For the high dose rate study,

a similar methodology was employed. Four couples of

PRESAGE cuvettes were placed within a slab in a solid

water phantom and irradiated at different dose-rates by

varying the dose delivery rate of an ELEKTA Versa HD FFF

linac from 2.5 up to 19 Gy/min. Dose delivery of 1 Gy for

all dose rates was verified by ion chamber measurements.

Irradiation induced optical density (OD) change was

measured from pre- and post-irradiation scans with a

digital spectrophotometer operated at 633 nm. Mean OD

change for each group was normalized to the value for the

highest dose rate in each study.

Results

Results presented in figure 1 show a trend of increasing

PRESAGE dose sensitivity with decreasing dose rate with

the over-response reaching up to 16% at 0.018 Gy/min.

Although in a first approach such low dose rates could be

considered extremely low in external radiotherapy, recent

studies have shown that in advanced radiotherapy

techniques (e.g. VMAT) dose rate varies drastically across

dose distributions delivered and a considerable

contribution of the delivered dose could come from very

low dose rates (0.01 - 0.1Gy/min). Regarding the high dose

rate study, all responses agree within experimental

uncertainties, indicating that PRESAGE sensitivity is not

significantly affected.

Figure 1: Dose rate dependence of PRESAGE response for

both studies included in this work. Error bars correspond

to 1 standard deviation of all experimental uncertainties

involved.

Conclusion

Results of this study indicate a significant over-response

of this PRESAGE formulation in very low dose rates that

should be considered when they are used in applications

involving wide range of dose delivery rates.

Acknowledgement: This work was financially supported by

the State Scholarships Foundation of Greece through the

program ‘Research Projects for Excellence IKY/SIEMENS’.

PO-0775 Contributions to detector response in

arbitrary photon fields

S. Wegener

1

, O.A. Sauer

1

1

University Hospital, Radiation Oncology, Würzburg,

Germany

Purpose or Objective

Due to their small active volumes, diodes are often the

detectors of choice for many commissioning tasks

including the measurement of output factors, especially in

small fields. However, high-atomic number material in the

chip, detector shielding or other components and a finite

active volume size have been found to alter the signal

compared to the dose ratios measured in water in the

absence of such a detector. As a consequence, correction

factors need to be applied to correct the obtained signals.

Using three experimental setups (fig. 1), the different

contributions to the detector signals were separated and

analyzed: the response to scatter, the primary beam and

the combination of both.

Material and Methods

Signal ratios were obtained for three different

experimental setups (fig. 1): First, the standard open field

geometry. Secondly, fields in which the central part of the

beam was blocked out by a 4 mm aluminum pole and the