Load capped tubes into a clean and decontaminated 3M Molecular Detection Speed Loader

3

Tray. Close and latch the 3M Molecular Detection Speed Loader Tray lid.

4

5

6

7

17.

Review and confirm the configured run in the 3M Molecular Detection Software.

8

18.

Click the Start button in the software and select instrument for use. The selected

9

instrument’s lid automatically opens.

10

19.

Place the 3M Molecular Detection Speed Loader Tray into the 3M Molecular Detection

11

Instrument and close the lid to start the assay. Results are provided within 75 minutes,

12

although positives may be detected sooner.

13

20.

After the assay is complete, remove the 3M Molecular Detection Speed Loader Tray from

14

the 3M Molecular Detection Instrument and dispose of the tubes by soaking in a 1-5% (v:v in

15

water) household bleach solution for 1 hour and away from the assay preparation area.

16

NOTICE: To minimize the risk of false positives due to cross-contamination, never open reagent

17

tubes containing amplified DNA. This includes Reagent Control, Reagent, and Matrix Control

18

tubes. Always dispose of sealed reagent tubes by soaking in a 1-5% (v:v in water) household

19

bleach solution for 1 hour and away from the assay preparation area.

20

21

R

ESULTS AND

I

NTERPRETATION

22

An algorithm interprets the light output curve resulting from the detection of the nucleic acid

23

amplification. Results are analyzed automatically by the software and are color-coded based on

24

the result. A Positive or Negative result is determined by analysis of a number of unique curve

25

parameters. Presumptive positive results are reported in real-time while Negative and Inspect

26

results will be displayed after the run is completed.

27

28