Dual Lab Validation of a Method for the Determination of Fructans in Infant
Formula & Adult Nutritionals /
25 Jul 2016
57
Results calibration test 10 May 2016
Fructose calibration ISC 5000_E instrument
Fructose
range: 0.887 – 177.4 mg/l
y = 0.1199 x
2
+ 13.574x + 0.03780
R
2
= 0.9999
Glucose calibration ISC 5000_E instrument
Glucose
range 0.054 – 21.78 mg/l
y = 0.052 x
2
+ 6.539 x +0.006 R
2
= 0.9999
respons
fructose
fructose
Fructose/chitobiose
conc
calculated
lack of fit
tov IS
mg/ml
absolute %
0,0159
0,2215
0,2373
‐0,016
‐7,1
0,0966
1,329
1,253
0,076
5,7
0,162
2,215
2,074
0,141
6,3
0,341
4,430
4,345
0,085
1,9
0,684
8,860
8,694
0,166
1,9
1,23
15,50
15,72
‐0,22
‐1,4
1,72
22,15
21,99
0,16
0,7
3,49
44,30
45,33
‐1,03
‐2,3
4,99
66,45
65,82
0,63
0,9
6,66
88,60
89,07
‐0,48
‐0,5
9,61
132,9
132,0
0,9
0,7
12,6
177,2
177,6
‐0,4
‐0,2
respons
glucose
fructose
glucose/chitobiose
conc
calculated
lack of fit
mg/l
mg/ml
absolute %
0,0080
0,0511
0,0504
0,0006
1,3
0,0168
0,1022
0,1006
0,0016
1,5
0,0347
0,2043
0,2019
0,0024
1,2
0,0572
0,3065
0,3300
‐0,0235
‐7,7
0,0853
0,5108
0,4898
0,0210
4,1
0,182
1,022
1,040
‐0,018
‐1,8
0,358
2,043
2,046
‐0,003
‐0,1
0,635
3,576
3,633
‐0,057
‐1,6
0,863
5,108
4,953
0,155
3,0
1,795
10,22
10,39
‐0,17
‐1,7
2,606
15,32
15,20
0,13
0,8
3,479
20,43
20,46
‐0,03
‐0,2
VALIDATION REPORT
FOR ERP USE ONLY
DO NOT DISTRIBUTE