Table of Contents Table of Contents
Previous Page  421 / 1195 Next Page
Information
Show Menu
Previous Page 421 / 1195 Next Page
Page Background

Page

19

of

38

For the low-level after a 24 hour incubation in RV10 broth, there were 4 presumptive and 4 confirmed positives for

the candidate method and 6 confirmed positives following the FDA/BAM reference method. A POD

C

value of 0.20

was obtained with 95% confidence intervals of (0.08, 0.42). A POD

R

value of 0.20 was obtained with 95%

confidence intervals of (0.08, 0.42). A dPOD

C

value of -0.10 was obtained with 95% confidence intervals of (-0.35,

0.17) indicating no statistical difference between the candidate and reference method. For the low inoculation level a

POD

CP

value and a POD

CC

value of 0.20 was obtained with 95% confidence intervals of (0.08, 0.42). A dPOD

CP

value of 0.00 with 95% confidence intervals of (-0.25, 0.25) indicating no statistical difference between the

candidate method presumptive and confirmed results

For the high inoculation level, there were 4 presumptive and 4 confirmed positives for the candidate method and 4

confirmed positives following the reference method. A POD

C

and POD

R

value of 0.80 was obtained with 95%

confidence intervals of (0.38, 1.00). A dPOD

C

value of 0.00 was obtained with 95% confidence intervals of (-0.47,

0.47), indicating no significant difference between the candidate and reference method. A POD

CP

value and a

POD

CC

value of 0.80 was obtained with 95% confidence intervals of (0.38, 1.00). A dPOD

C

value of 0.00 was

obtained with 95% confidence intervals of (-0.47, 0.47), indicating no statistical difference between the candidate

method presumptive and confirmed results.

Stainless Steel

For the low-level, there were 10 presumptive and 10 confirmed positives for the candidate method and 9 confirmed

positives following the FDA-BAM reference method. A POD

C

value of 0.50 was obtained with 95% confidence

intervals of (0.30, 0.70). A POD

R

value of 0.45 was obtained with 95% confidence intervals of (0.26, 0.66). A

dPOD

C

value of 0.05 was obtained with 95% confidence intervals of (-0.24, 0.33) indicating no statistical difference

between the candidate and reference method. For the low inoculation level a POD

CP

value and a POD

CC

value of

0.50 was obtained with 95% confidence intervals of (0.30, 0.70). A dPOD

CP

value of 0.00 with 95% confidence

intervals of (-0.28, 0.28) indicating no statistical difference between the candidate method presumptive and

confirmed results.

For the high inoculation level, there were 5 presumptive and 5 confirmed positives for the candidate method and 5

confirmed positives following the reference method. A POD

C

and POD

R

value of 1.00 was obtained with 95%

confidence intervals of (0.57, 1.00). A dPOD

C

value of 0.00 was obtained with 95% confidence intervals of (-0.43,

0.43), indicating no statistical difference between the candidate and reference method. A POD

CP

value and a POD

CC

value of 1.00 was obtained with 95% confidence intervals of (0.57, 1.00). A dPOD

C

value of 0.00 was obtained with

95% confidence intervals of (-0.43, 0.43), indicating no statistical difference between the candidate method

presumptive and confirmed results.

Dry Pet Food-25 g

For the low-level, there were 5 presumptive and 5 confirmed positives for the candidate method and 4 confirmed

positives following the FDA/BAM reference method. A POD

C

value of 0.25 was obtained with 95% confidence

intervals of (0.11, 0.47). A POD

R

value of 0.20 was obtained with 95% confidence intervals of (0.08, 0.42). A

dPOD

C

value of 0.50 was obtained with 95% confidence intervals of (-0.21, 0.30) indicating no statistical difference

between the candidate and reference method. For the low inoculation level a POD

CP

value and a POD

CC

value of

0.25 was obtained with 95% confidence intervals of (0.11, 0.47). A dPOD

CP

value of 0.00 with 95% confidence

intervals of (-0.26, 0.26) indicating no statistical difference between the candidate method presumptive and

confirmed results.

For the high inoculation level, there were 4 presumptive and 4 confirmed positives for the candidate method and 4

confirmed positives following the reference method. A POD

C

value of 0.80 was obtained with 95% confidence

intervals of (0.38, 1.00) and POD

R

value of 0.80 was obtained with 95% confidence intervals of (0.38, 1.00). A

dPOD

C

value of 0.00 was obtained with 95% confidence intervals of (-0.47, 0.47), indicating no statistical difference

between the candidate and reference method. A POD

CP

value and a POD

CC

value of 0.80 was obtained with 95%

confidence intervals of (0.38, 1.00). A dPOD

C

value of 0.00 was obtained with 95% confidence intervals of (-0.47,

0.47), indicating no statistical difference between the candidate method presumptive and confirmed results.

Discussion

2013.09 /PTM Matrix Extension Report

March 2014

Expert Review Panel Use Only

AOAC Research Institute

Expert Review Panel Use Only