automated sample preparation and automated PCR preparation to minimize contamination.
The
mericon
kitsallow the end user a choice of two DNA extraction methods. The
automated
QIAsymphony
preparation method allows a high volume of samples to be
processed with minimal supervision utilizing easy to use software with an option to follow
step-by-step guidance through the sample prep workflow.
Conclusion
The data from these studies, within their statistical uncertainty, support the product claims
of the
mericon
E. coli O157Screen Plus and
mericon
E. coliSTEC O-TypePathogen
Detection Assaysfor the selected food matrixes tested, namely fresh spinach, raw ground
beef, and raw beef trim.
Results obtained from POD analysis in the method comparison study demonstrated no
statistically significant differences between the number of positive samples detected by the
candidate method, using the manual or automated DNA extraction procedures, and the
reference methods for fresh spinach, raw ground beef, and raw beef trim.
OMAMAN-36 E/ AOAC PTM Report
ERP Use Only
January 2017
AOAC Research Institute
Expert Review Panel Use Only