![Show Menu](styles/mobile-menu.png)
![Page Background](./../common/page-substrates/page0370.png)
Page
43
of
76
Figure 7. Stability
mericon+
E. coli O157 Screen Plus assay
There was no decrease (right shift)observed between the time points t=0 @ 4
+ 1
o
C, t=4 months @ 4 + 1
o
C or t+4 months @ 37 + 1
o
C (about 18 months
equivalent storage). Storing the prime-probe lots had, therefore, no
deleterious effect on the performance of the assay.
Figure 8. Stability
merion
E. coli STEC O-type assay
0.00
5.00
10.00
15.00
20.00
25.00
30.00
35.00
t=0
4 months 4 months @37
Ct
Storage, months
Stability mericon
E. coli Screen Plus:
eae
10000
1000
100
25
10
0.00 5.00 10.00 15.00 20.00 25.00 30.00 35.00
0 months
4 months
Ct
Storage, months
mericon E. coli O-type assay: Big 6
Stability
10000 1000 100 25 10
OMAMAN-36 E/ AOAC PTM Report
ERP Use Only
January 2017
AOAC Research Institute
Expert Review Panel Use Only