Figure 6

.mericonE. coli

STEC O-type assay: Prep lot variation

Figure 6.

mericon

E. coli STEC O-type Prep lot variation

Ct values for DNA prep lots using consistent primer probe:master mix lots for

the

mericonE. coli

STEC O-type Pathogen Detection Assaywere not

significantly different, indicating that there were no effects of different

bacterial prep lots on the detection results

Stability Study: Lot stability

In this study, a single lot of each of the primer probe assay kits (

E. coli

O157Screen Plusand

E. coil

STEC O-type)was stored at 4 + 1

o

C for 4 and

15 months. A second set of kits from the same lot was stored at 37 + 1

o

C for

3 and 4.5 months. The 2 sets of primer probes stored at 37 + 1

o

C extends the

0.00 5.00 10.00 15.00 20.00 25.00 30.00

E.coli

O157:H7

E.coli O157 E.coli

O26:H-

E.coli

O111:H-

E.coli O145 Citrobacter

freundii

Ct

Bacterial Strain

mericon

E. coli O-type O157:H7: Lot-to-Lot prep

Prep 1

Prep 2

Prep 3

0 5 10 15 20 25 30

E.coli

O157:H7

E.coli O157 E.coli O26:H-

E.coli

O111:H-

E.coli O145 Citrobacter

freundii

Ct

Bacterial Strain

mericon

E. coli O-type Big 6: Lot-to-Lot prep

Prep 1

Prep 2

Prep 3

OMAMAN-36 E/ AOAC PTM Report